Descripción

Withstanding Voltage Tester

Características
High power: AC 5kV / 100mamá / 500VA output
High security:
High-voltage floating output design, in line with the safety
requirements of EU standards EN50191 (only TH9110)
Electric shock protection function
High resolution: 7 pulgada 800 × 480 puntos, TFT-LCD display
Brand-new operation interface, Menú chino e inglés.
ARC detection function
Función de verificación de contacto (OSC)
Breakdown voltage test function
One-key screen capture function
One-key recording function
Rear panel output function to facilitate automated production line testing
Almacenamiento: 100 archivos, hasta 50 steps per file

Withstanding Voltage Tester

Solicitud
Winding devices
Transformers, generators/motors and other products needing high
-power withstand voltage test and analysis, such as different types of
motor stator, rotor and other high parasitic capacitance products
Electronic components
Capacitors, coils, cores, choke coils, filters and so on
Electrical products
Household appliances, information products, audio-visual
equipo, aparatos de calefacción eléctrica, lighting equipment

Withstanding Voltage Tester

Parámetro técnico

Modelo TH9110 TH9110A
Withstand Voltage Test
Tensión de salida C.A. 0.05 – 5kV Load Variance:1% Exactitud:1% Resolución: 2V
corriente continua 0.05 – 6kV Load Variance:1% Exactitud:1% Resolución: 2V
Current Test Range C.A. 0.001mamá – 120mamá(Voltage≤4kV);0.001mamá – 100mamá(Voltaje>4kV) Exactitud:1% Resolución:1µA
corriente continua 0.0001mamá – 25mA Accuracy:1% Resolución:0.1 µA
Output Power 500Virginia
Insulation Resistance Test
Tensión de salida : 0.05 – 5kV Resolution:2V Accuracy:1% of set value + 0.1% escala completa
Rango de prueba de resistencia 1MΩ-50.0GΩ Resolution:0.1MΩ
Discharge Function Automatic discharge after the end of the test
Detección de arco C.A. 1mamá – 20mamá
corriente continua 1mamá – 10mamá
Contact Check Function OSC open and short:600Hz, 0.1s
Security Features
High Voltage Floating Output Leakage current <3 mamá ——
Electric Shock Protection 0.5mA ±0.25mA
Other Protection Start protection, panel operation password protection
Alarm Indication APROBAR: short tone, green light; FAIL: long tone, red light
Memoria grupos, 50 steps per group
General Parameters
Voltage Rise Time 0.1s — 999.9s
Test Time Setting(AC/DC) 0.3s — 999s
Voltage Fall Time 0.1s — 999.9s
Waiting Time (Y) 0.2s — 999.9s
Time Accuracy ±(1%+0.1s)