Máy kiểm tra phân bố kích thước hạt FO9424

Đặc trưng:

The percentage of the total amount of particles in different particle size intervals determined by a specific method is called the particle size distribution, and some users also call it the particle size distribution. Particle size distribution (particle size distribution) has a variety of benchmarks, such as quantity distribution, length distribution, surface area distribution, volume distribution, vân vân., different units of measurement will cause great changes in the form of particle size distribution (particle size distribution) results.

The size distribution is usually given in the form of a size distribution table or/and a size distribution curve. (1) Table form: All particle size intervals and their corresponding percentage of content are listed in the table, with differential distribution and cumulative distribution in two forms. (2) Curve form: Particle size distribution curve is an image expression of particle size distribution. Compared with the grain size distribution table, the curve has the advantages of image, intuition and clear glance. The particle size distribution curve is also divided into differential distribution curve and cumulative distribution curve, which has the same physical meaning as the particle size distribution table.

 

Máy kiểm tra phân bố kích thước hạt FO9424

Determination procedure
1. Turn on the power supply of the particle size analyzer (hereinafter referred to as the instrument) and preheat for 15min.
2. According to the use procedure of the instrument, enter the true density of dust, the height of the liquid level, and the density and viscosity of the acetate at the measured temperature.
3. After the instrument prints out the input value, pour butyl acetate into the clean instrument measuring tank, and pour the amount of about 2/3 of the liquid level. Put the measuring tank containing butyl acetate into the measuring position of the instrument and adjust the zero control knob so that the optical density value displayed by the instrument is 0.0.
4. Put the dust-containing filter film into a clean porcelain crucible or beaker, drop 5~10mL butyl acetate and stir thoroughly with a glass rod. When stirring, the dust-containing liquid is dropped into the measuring tank containing butyl acetate with pipette, so that the liquid level reaches the input liquid level height scale line.
5. Replace butyl acetate with anhydrous ethanol for coal mine dust that is generally not taken on the filter membrane. The method is to take 3 ~ 5g dust on the white paper fully mixed evenly, with a horn spoon to take about 10mg of dust into a clean crucible or beaker, in accordance with the method described in Article 4.
6. Cover the measuring pool cover, hold the measuring pool tightly with your thumb and shake the cover fully, so that the dust in the measuring pool is evenly dispersed, and dry the liquid on the surface of the measuring pool with absorbent cotton gauze. Quickly put the measuring tank into the measuring position of the instrument, so that the optical density value displayed by the instrument is within the range of 90 ĐẾN 100, and quickly press the measuring key to determine the dust powder distribution. If the optical density value initially displayed by the instrument is not in the range of 90 ĐẾN 100, it means that the concentration of the suspension is too high or too low, and the dust concentration in the liquid needs to be re-adjusted.
7. After pressing the measurement key, the instrument begins to display the measurement time and the optical density value of the corresponding time at an interval of 0.5s. Press the following formula to calculate

 

Máy kiểm tra phân bố kích thước hạt FO9424

Tham số

Mục index
nguyên tắc đo lường nhiễu xạ laser
mô hình quang học Lý thuyết Mistral đầy đủ và lý thuyết Fraunhofer là tùy chọn
phạm vi kích thước hạt 0.1μm-1200μm, không cần thay ống kính, không dựa vào việc hiệu chuẩn mẫu chuẩn
hệ thống phát hiện Includes a grille type large Angle, mảng dò tìm bù diện tích chéo không đồng nhất
tế bào đo Độ nghiêng song song
nguồn sáng 638Laser trạng thái rắn nm với hệ thống ổn nhiệt tích hợp, lên tới 20mW
Chế độ lọc không gian Công nghệ lọc phân cực không lỗ kim
Hệ thống căn chỉnh quang học Thông minh hoàn toàn tự động
đo thời gian Giá trị điển hình nhỏ hơn 10 giây
đo tốc độ Lên đến 20,000 lần mỗi giây
mức độ chính xác Dv50 tốt hơn ±0,6% (NIST-Tiêu chuẩn latex có thể truy nguyên)
mức độ chính xác Dv50 tốt hơn ±0,5% (NIST-Tiêu chuẩn latex có thể truy nguyên)
tia laser an toàn Lớp học 1 sản phẩm laser
cấu hình máy tính Bộ xử lý Intel i5,4RAM GB,250ổ cứng GB, chuột, bàn phím và màn hình rộng
giao diện máy tính USB2.0 trở lên
Nền tảng chạy phần mềm Windows7 trở lên Phiên bản chuyên nghiệp
vận hành Nhiệt độ xung quanh 5oC-40oC
Độ ẩm môi trường xung quanh hoạt động 10%-85% độ ẩm tương đối (không ngưng tụ)
yêu cầu năng lượng AC 220V, 50Hz-60Hz, mặt đất tiêu chuẩn
Trọng lượng hệ thống quang học 30Kilôgam
Kích thước hệ thống quang học 670mmx275mmx320mm