CLD Ceramic Antifreeze Tester
Introduction
The instrument is suitable for measuring the frost resistance of ceramic brick special equipment, the instrument meets the requirements of GB/3810.12-2016 and ISO10545-12:1995 ceramic brick frost resistance.
Main technical parameters
1, the test area operating temperature adjustment range: -10℃ – +10℃;
2, test area: 600×600×300 (mm);
3, temperature control accuracy: ±1℃;
5, test mode: according to the set number of tests automatically completed cycle.
6, power supply: 220V ±10% 2.5kw;
Working conditions
1, ambient temperature: 0 ~ 40℃;
2, power supply: 220V, 50 HZ;
3, equipped with inlet water source, outlet pipe;
CLD Ceramic Antifreeze Tester
instrument composition
The instrument consists of three parts: freezing test chamber, inlet and drainage system, temperature control and automatic control system.
The working principle of the instrument is designed and realized according to the requirements of GB/3810.12-2016 national standard.
Set the number of test cycles (100 times) for the electronic counter according to the test requirements. Put the sample into the test area and cover the cover. Refrigerate the test area to -5˚C for 15 minutes, turn on the inlet pump to spray water on the sample to increase the test temperature to +5˚C, hold for 15 minutes, count the counter once, set the electronic counter according to the test requirements for the number of test cycles (100 times). Automatically according to the above cycle test, after the test alarm notice.