Description

GA-G138 Portable Infrared Flue Gas Analyzer
Présentation du produit:

The portable infrared flue gas analyzer utilizes non-dispersive infrared gas sensors and electrochemical gas sensors, equipped with an imported integrated sampling and heating device and an imported portable pretreatment device. It can simultaneously measure the volume concentration of gases such as SO2, NON, CO, CO2, and O2 in flue gas, as well as parameters such as flue gas temperature and flow rate, and calculate emission rates and total emissions. It is a high-performance portable product specifically designed for environmental monitoring, energy conservation monitoring, energy efficiency testing, and environmental science research.

GA-G138 Portable Infrared Flue Gas Analyzer
caractéristiques du produit:

Précision de mesure élevée. Utilizes proprietary non-dispersive infrared gas analysis technology and long-life electrochemical sensing technology, ensuring no cross-interference between multi-component measured gases and high measurement accuracy.
High data portability. Automatically stores measurement data and can automatically calculate excess air coefficient and combustion efficiency. Equipped with a dot matrix printer and USB communication interface, it can print real-time and historical stored data, or directly export stored data to a USB flash drive.
Powerful measurement parameter expansion capabilities. Can simultaneously measure the volume concentration of multiple gases in flue gas, and can be expanded to measure flue gas temperature, pressure, flow rate, température ambiante, and pressure.
High-performance sampling and pretreatment device. The analyzer is equipped with an original imported integrated sampling and heating device and an original imported aluminum alloy portable pretreatment device to ensure that the sample gas meets the instrument requirements.

GA-G138 Portable Infrared Flue Gas Analyzer
Application Areas:

Environmental monitoring and enforcement, boiler and kiln efficiency testing, checking burner operating conditions, checking equipment leaks

GA-G138 Portable Infrared Flue Gas Analyzer
Paramètres techniques
Measured Components SO2、NON、CO、CO2、Ô2
Measurement Principle SO2/NO/CO/CO2:NDIR;O2:DPE
Plage de mesure SO2/NO:(0~2000)ppm
CO:(0~5000)ppm
CO2/O2:(0~25)%
Linear Error SO2/NO/CO/CO2:±2%F.S.;O2:±2%F.S.
Résolution SO2/NO/CO:1ppm;CO2/O2:0.01%
Répétabilité ≤2%
Flue Gas Temperature (0~800)℃
Flue Gas Velocity (3~40)MS
Response Time T90<15s(NDIR)
Optimal Flow Rate (0.7~1.2)L/min
Inlet Pressure (2~50)kPa
Communication USB, dot matrix printer
Source de courant 100~240VAC,150W
Functional Configuration
Equipped with self-diagnostic function for online sensor status checking
Heated sampling probe with built-in filter element
Corrosion-resistant sampling probe, freely combinable into different lengths such as 1m, 2m, etc., to adapt to different site requirements
Imported self-regulating heated tubing, heating temperature 180℃
Equipped with a pre-treatment device, including a built-in sampling pump, condenser, and peristaltic pump
Expandable to include flue gas temperature, flue gas pressure, flue gas velocity, température ambiante, and ambient pressure