Description

Mesureur d'épaisseur de revêtement GA-3816

Introduction

The Times GA-3816 thickness gauge is a portable measuring instrument, which can measure the thickness of coating and coating rapidly and precisely without damage.It can be used in both laboratory and engineering field.Through the use of different probe, can meet the needs of a variety of measurement.Scope of era GA-3816 thickness gauge using both the magnetic and eddy current thickness method, measuring magnetic metal substrate condition (such as steel, fer, alloy and hard magnetic steel, etc.) on the magnetic coating thickness (such as aluminum, chrome, cuivre, enamel, etc.), caoutchouc, paint and nonmagnetic metal substrate (comme le cuivre, aluminium, zinc, étain, etc.) on the conductive layer thickness.

 

Mesureur d'épaisseur de revêtement GA-3816

Character and function

Using both magnetic and eddy current thickness method, and F type measuring head without damage to the measurement of magnetic metal basic body of magnetic coating thickness (such as steel, fer, zinc alloy and hard magnetic steel, aluminium, chrome, cuivre, caoutchouc, peinture, etc.), and N type measuring head without damage to the measurement of non-magnetic thickness of the conductive coating on metal substrates (comme le cuivre, aluminium, zinc, tin and other metal oxide film of paint, caoutchouc, Plastique, etc.)

Can use 6 kinds of measuring head / 90 ° F400, F1, F1, F10, N1, CN02

Two measurement methods: CONTINUE and SINGLE

Two ways of working: DIRECT and group (APPL)

There are five statistics: MEAN, larger (MAX), MIN, number of tests (NO.), and standard deviation (S.DEV).

Two methods can be used to calibrate the instrument

Chinese and English menu display, operation more intuitive and convenient

Liquid crystal display has backlighting function

It can store 640 données

With delete function, delete single or all data measured

Print function

With integrated printer, facile à utiliser

Power supply undervoltage indication

The operation process is beeping (single measurement method)

Error message function

Manual and automatic shutdown functions

 

Mesureur d'épaisseur de revêtement GA-3816

TT270 thickness gauge head selection table is as follows:

Probe model F400 F1 F1/90° F10
Work Theory Magnetic induction
Test range(um) 0~400 0~1250 0~10 000
Low resolution(um) 0.1 0.1 10
Value error One point calibration(um) ±(3%H+1) ±(3%H+10)
Two point calibration(um) ±[(1~3) %H+0.7] ±[(1~3)%H+1] ±[(1~3)%H+10]
 

Test

conditions

 

Minimum radius of curvature(um) convex 1 1.5 flat 10
The diameter of the smallest area(mm) 3 7 7 40
Critical matrix thickness(mm) 0.2 0.5 0.5 2
Probe model N1 CN02
working principle Eddy current
Measuring range(um) 0~1250 10~200
Low resolution(um) 0.1 1
Value error One point calibration(um) ±(3%H+1.5) ±(3%H+1)
Two point calibration(um) ±[(1~3) %H+1.7] —
test

conditions

 

Minimum radius of curvature(um) 3 Only for flat
The diameter of the smallest area(mm) 5 7
Critical matrix thickness(mm) 0.3 unlimited