Aparato automático de conductividad térmica DRH-II-300
- Descripción
- Consulta
Descripción
Aparato automático de conductividad térmica DRH-II-300
Introducción
The instrument is based on the principle of unidirectional stable heat conduction. When the upper and lower sides of the sample are at different stable temperatures, the heat flow through the effective heat transfer area of the sample and the temperature difference and thickness between the two surfaces of the sample are measured to calculate the thermal conductivity. It meets the high precision test requirements of material testing research department for thermal conductivity of materials. Mainly test plastic, vaso, fibra, foam, insulation materials, etc.. Widely used in colleges and universities, scientific research units, quality inspection departments and production plant material analysis and testing.
Estándar de referencia del instrumento:
GB/10294-2008 (Determination of steady-state thermal resistance and related properties of thermal insulation materials protective hot plate method);
GB/T3399-1982 (test method for thermal conductivity of plastics, heat shield plate method);
GB/T3139-2005 (test method for thermal conductivity of fiber reinforced plastics) (test method for thermal conductivity of glass fiber reinforced plastics).
GB/T10801.1-2002 (polystyrene foam for thermal insulation) requirements for measurement of thermal conductivity of polyurethane rigid foam materials;
GB/T10801.2-2002 (Extruded polystyrene foams for thermal insulation (XPS)) – Requirements for measurement of thermal conductivity of extruded polystyrene foams;
GB/T3139-2005 (Test method for thermal conductivity of fiber reinforced plastics);
GB/T 17794-2008 (productos aislantes de gomaespuma flexible) requirements for the measurement of thermal conductivity.
Características de presentación:
1, automatic thickness measurement: the use of high-precision displacement sensor measurement, and connected to the computer automatic thickness measurement, reducir el error humano.
2, automatic pressure system: computer automatic pressure, can test the thermal conductivity under different pressures, more scientific than manual pressure, más datos, better repeatability.
3, the use of double flat detection, higher accuracy.
Principales indicadores técnicos:
1, rango de conductividad térmica: 0.001 ~ 3.000W/mk;
2, precisión de la prueba: ± 3%;
3, tamaño de la muestra: double specimen 300×300×(10-50)milímetros; Flatness: < 0.05 milímetros.
4, rango de temperatura de la superficie caliente: room temperature ~ 100℃, resolución de temperatura 0,01 ℃, precisión de la medición: 0.05℃;
5, rango de temperatura de la superficie fría: 0 ~ 60℃, resolución de temperatura 0,01 ℃;
6, computer control automatic test, procesamiento de datos, and generate test reports, can be printed out;
7, the use of high-precision displacement sensor and connected to the computer automatic thickness measurement, measuring range 0– 50milímetros, resolution 0.01mm;
8, can be automatically pressurized, detect the thermal conductivity under different pressure states, pressure range 0– 1000N adjustable, resolution 0.1N.
9, hot and cold plate temperature difference: the default value of 20℃, users can also set their own
Common standard temperature difference: 20℃ Common temperature difference range 15-25℃
Configuración del instrumento:
1, DRH-II-300 automatic test host,
2, a set of software,
3, a high precision constant temperature sink,
4, una computadora de escritorio (según la configuración del contrato).