CLD Ceramic Antifreeze Tester
Introducción
The instrument is suitable for measuring the frost resistance of ceramic brick special equipment, the instrument meets the requirements of GB/3810.12-2016 and ISO10545-12:1995 ceramic brick frost resistance.
Principales parámetros técnicos.
1, the test area operating temperature adjustment range: -10℃ – +10℃;
2, test area: 600×600×300 (milímetros);
3, precisión del control de temperatura: ±1℃;
5, Modo de prueba: according to the set number of tests automatically completed cycle.
6, fuente de alimentación: 220V ±10% 2.5kw;
Las condiciones de trabajo
1, temperatura ambiente: 0 ~ 40 ℃;
2, fuente de alimentación: 220V, 50 HZ;
3, equipped with inlet water source, outlet pipe;
CLD Ceramic Antifreeze Tester
instrument composition
The instrument consists of three parts: freezing test chamber, inlet and drainage system, temperature control and automatic control system.
The working principle of the instrument is designed and realized according to the requirements of GB/3810.12-2016 national standard.
Set the number of test cycles (100 veces) for the electronic counter according to the test requirements. Put the sample into the test area and cover the cover. Refrigerate the test area to -5˚C for 15 minutos, turn on the inlet pump to spray water on the sample to increase the test temperature to +5˚C, hold for 15 minutos, count the counter once, set the electronic counter according to the test requirements for the number of test cycles (100 veces). Automatically according to the above cycle test, after the test alarm notice.