STD-A介质损耗计
介绍
Dielectric loss and dielectric constant are important physical properties of various ceramic such as electro ceramics,ceramic for mounting purposes,capacitor etc. The instrument is used in determining dielectric loss angle tangent tan δ and dielectric constant (e) to know more about various factors that effect the dielectric loss and dielectric constant, and to provide basis for improving the properties of materials. It is widely used in application study of properties of inorganic nonmetal new materials of institutions of higher education and research institutions. The instrument is in line with the standard of GB/T5594.4-1985.
STD-A介质损耗计
Main Technical Index:
1. The Q value measurement
A. Q value measuring range: 5 〜 999.
乙. Q value range step: 30, 100, 300, 999, automatic shift.
C. the nominal error
频率范围: 25 kHz ~ 10 兆赫兹;
Inherent error: ≤5%±the full value of 2%;Work of error: ≤7%±the full value of 2%;
频率范围: 10 MHz ~ 50 兆赫兹;
Inherent error: ≤7%±the full value of 2%;Work of error: ≤10%±the full value of 2%;
2. The inductance measurement
A. 测量范围: 0.1 mu H ~ 1 H.
乙. grade divide: seven ranges.
0.1 〜 1 u H, 1 〜 10 u H, 10 〜 100 u H,
0.1 ~ 1mH, 1 〜 10 mH, 10 〜 100 mH, 100mH ~ 1 H.
3. The capacitance measurement
A. 测量范围: 1 〜 460 pF (above 460 pF capacitance measurement see using rules);
乙. the adjusting range capacitance
Main tuning capacitor: 40 〜 500 pF;
准确性: 以下 150 pF ±1.5pF; above 150 pF ±1%;
Fine-tuning capacitance: – 3 pF~ 0 〜 + 3 pF;
准确性: ± 0.2 pF.
4. The oscillation frequency
A. oscillation frequency range: 25 kHz ~ 50 兆赫兹;
乙. the frequency step:
10 〜 99 千赫; 100 〜 999 千赫;1.0 〜 9.99 兆赫兹;10 〜 50 兆赫兹.
C. frequency error: 2 X 10-4 ± 1 数字.
5. Q qualified preset functions indicated, preset range: 5 〜 999.
6. Equipment normal working conditions:
A) 温度: 0 ℃ ~ + 40 ℃;
乙) 相对湿度: 80%;
C) 电源: 220 V±22V, 50 Hz ±2.5 Hz