描述

ESD30K Electrostatic Simulation Generator

Product Overviews

Electrostatic discharge (ESD) is a common natural phenomenon. The effects of ESD on electrical and electronic equipments, devices or systems are everywhere. It is a very high degree of harmful electromagnetic energy. Only to improve the level of anti-static interference can ensure the safe use of electronic products. ESD generator is used to evaluate the electrostatic immunity performance of electrical components and electronic equipments by simulating real electrostatic discharge as defined in the standard IEC/ EN 61000-4-2. Kindly find below features & technical data of our model ESD 30K.

ESD30K Electrostatic Simulation Generator
特征
  • User friendly 7” Touch Panel Display
  • Support multi-language & facilitate users
  • Built-in environment self-test program
  • Programmable operation
  • Pre-programmed IEC/ EN 61000-4-2 test settings
  • Very convenient to replace the discharge modules to meet different standards of the test requirements
  • RS232 / USB Port, PC control operation & printable test report
ESD30K Electrostatic Simulation Generator
技术数据
模型 ESD30K
标准 IEC61000-4-2,GB/T 17626.2
Operation 7” Touch Panel Display
输出电压 0.2~±30kV
Polarity Positive/ negative/ alternating
Energy storage capacity 150pF
Discharge resistor 330哦
Discharge current rise time 0.6~1ns
The first peak current (±15%) 7.5A at 2kV/ 15.0A at 4kV/ 22.5A at 6kV/ 30.0A at 8kV
Current at 30ns (±30%) 4.0A at 2kV/ 8.0A at 4kV/ 12.0A at 6kV/ 16.0A at 8kV
Current at 60ns (±30%) 2.0A at 2kV/ 4.0A at 4kV/ 6.0A at 6kV/ 8.0A at 8kV
Holding Time (air discharge) ≥5s
Discharge Modes Contact Discharge & Air Discharge
Operating Modes Single discharge & repetitive discharges
Discharge Trigger 手动的, automatic or remote
Working Mode IEC level mode/ user mode/ program mode/ voltage rising patterns
Discharge Interval 0.05~9.99s
Discharge Count 1~9999
电源 AC220V ± 10%,50/60Hz
环境温度 15℃~35℃

 

Environment
  • Indoor use
  • Altitude not exceeding 1000metres
  • Ambient temperature 15℃~35℃
  • Relative humidity no more than 85%
  • No conductive dust, no fire or explosion hazard, no corroding metal or insulating gas, sine wave voltage waveform, waveform distortion rate≤5%
  • Earthing resistance not more than 0.5Ω