- 描述
- 询问
描述
DRH-II-300全自动导热仪
介绍
The instrument is based on the principle of unidirectional stable heat conduction. When the upper and lower sides of the sample are at different stable temperatures, the heat flow through the effective heat transfer area of the sample and the temperature difference and thickness between the two surfaces of the sample are measured to calculate the thermal conductivity. It meets the high precision test requirements of material testing research department for thermal conductivity of materials. 主要测试塑料, 玻璃, 纤维, 泡沫, insulation materials, ETC. 广泛应用于高等院校, 科研单位, quality inspection departments and production plant material analysis and testing.
仪器参考标准:
GB/10294-2008 (保温材料稳态热阻及相关性能的测定防护热板法);
GB/T3399-1982 (test method for thermal conductivity of plastics, heat shield plate method);
GB/T3139-2005 (纤维增强塑料导热系数试验方法) (玻璃纤维增强塑料导热系数试验方法).
GB/T10801.1-2002 (polystyrene foam for thermal insulation) requirements for measurement of thermal conductivity of polyurethane rigid foam materials;
GB/T10801.2-2002 (Extruded polystyrene foams for thermal insulation (XPS)) – Requirements for measurement of thermal conductivity of extruded polystyrene foams;
GB/T3139-2005 (Test method for thermal conductivity of fiber reinforced plastics);
国标/T 17794-2008 (flexible foam rubber insulation products) requirements for the measurement of thermal conductivity.
性能特点:
1, automatic thickness measurement: the use of high-precision displacement sensor measurement, and connected to the computer automatic thickness measurement, reduce human error.
2, automatic pressure system: computer automatic pressure, can test the thermal conductivity under different pressures, more scientific than manual pressure, more data, better repeatability.
3, the use of double flat detection, 更高的精度.
Main technical indicators:
1, 导热系数范围: 0.001 ~ 3.000W/mk;
2, 测试准确度: ± 3%;
3, 样本量: double specimen 300×300×(10-50)毫米; Flatness: < 0.05 毫米.
4, 热面温度范围: room temperature ~ 100℃, temperature resolution 0.01℃, measurement accuracy: 0.05℃;
5, 冷面温度范围: 0 ~60℃, temperature resolution 0.01℃;
6, computer control automatic test, 数据处理, 并生成测试报告, can be printed out;
7, the use of high-precision displacement sensor and connected to the computer automatic thickness measurement, 测量范围 0– 50毫米, 分辨率0.01mm;
8, 可自动加压, 检测不同压力状态下的导热系数, 压力范围0– 1000N可调, 分辨率0.1N.
9, hot and cold plate temperature difference: the default value of 20℃, users can also set their own
Common standard temperature difference: 20℃ Common temperature difference range 15-25℃
Instrument configuration:
1, DRH-II-300 automatic test host,
2, a set of software,
3, a high precision constant temperature sink,
4, 一台台式计算机 (根据合同配置).