X-ray fluorescence spectrometer FAQ Series 1
Q: What are the six substances restricted by RoHS?
A: Hexavalent chromium, cadmium, lead, mercury, polybrominated biphenyls and polybrominated diphenyl ethers.
Q: What are the maximum content limits for each of the six substances restricted by RoHS?
A: The hexavalent is 1000PPM, cadmium is 100PPM, mercury is 1000PPM, lead is 1000PPM, polybrominated biphenyls are 1000PPM, and polybrominated biphenyls are 1000PPM.
Q: What is the range of elements that EDX9000B can detect?
A: It can detect elements from sodium to uranium.
Q: What kind of tests can our equipment do?
A: It can do RoHS testing, halogen (Br and CI) testing, toy eight heavy metal testing, full element analysis of various materials and measurement of metal coating thickness.
Q: How does EDX equipment work?
A: Principle: The primary X-ray is produced by the flow of electrons generated at high pressure to reach the target material in the X-ray tube. The primary X-ray is filtered and concentrated and shot into the sample under test to produce secondary X-ray, which is usually called X-ray fluorescence. After being detected by the X-ray fluorescence backup detector, it is amplified and input into the computer through digital-analog conversion.
Q: Is our equipment imported or made in China?
A: Our equipment is made in China, but the main parts are imported, such as detector and light tube from America.
X-ray fluorescence spectrometer FAQ Series 1
Q: Does EDX equipment cause harm to the human body and pollution to the environment?
A: Our company’s products have been certified by the National Environmental Radiation Research and Monitoring Center, and the radiation is far lower than the national 2500ngy standard. At the same time, the instrument has triple ray protection function, which will not cause any harm to the human body, nor will it cause direct or indirect pollution to the environment.
Q: Has our equipment passed the relevant safety and environmental protection certification?
A: X-ray fluorescence analysis equipment is not subject to strong inspection. But our equipment has been certified by the relevant departments of the National Environmental Radiation Research and Monitoring Center.
Q: Is the instrument complicated to use and operate?
A: No. For different industry applications, we have different software applications, adapt to the requirements of each industry, ordinary operators as long as through our simple training can skillfully operate
Q: Can you measure the coating?
A: Yes. Samples of single, double and multiple coatings; Moreover, all coating thickness is tested in one measurement, the test speed is fast, and the test results are accurate and convenient
Q: Is the software available in English and traditional?
A: There are both English and traditional versions. Our software is mainly Chinese, English and traditional three, for RoHS software it has a third-party language editing function, can be edited by customers according to their own requirements into any national language.
Q: Why does the background appear?
A: When the X-ray fluorescence analysis instrument is tested, there will be scattering, free electrons, line interference, etc., which will cause background, which can be commonly known as “background noise”.
Q: What is our basis for elemental analysis?
A: Characteristic X-rays, which are determined by the basic constituent elements of the substance being measured. Different elements have different characteristic X-ray energies.
Q: What are the advantages of the instrument in the analysis and testing of the hardware industry and the steel industry?
A: Fast, accurate, non-destructive sample, simple pretreatment, easy to operate.
Q: X-ray fluorescence tester in heavy metals, petroleum exploration industry application examples explain.
A: Please refer to the company’s publicity materials.
Q: How do some factories and laboratories test elements? Good and bad?
A: There are many test methods, and used in different products and industries, the test methods are not the same, the advantages of each instrument is also different, which instrument is good, but also depends on the customer’s real application field and the actual test sample. 1
X-ray fluorescence spectrometer FAQ Series 1
Q: Can you provide test report and relevant certificate after instrument testing?
A: The test report can be provided, but the certificate of the report cannot be provided. Because certification is for the testing organization, it is not for the instrument. The report of any X-ray fluorescence spectrometer in the world is not authoritative.
Q: Can the EDX800B test a finished product?
A: It can test a complete finished product. According to the test method of EU and IEC, it is for homogeneous substances, and the finished product must be physically split to the point where it cannot be split, and then tested.
Q: If you buy your instrument, can you guarantee the report of the test on the product in use?
A: No, because the X-ray fluorescence spectrometer is only a crude measuring equipment, strictly speaking, it is semi-accurate quantitative, its measurement results do not have authority, at the same time, your company’s inspector’s operation is normal, are the factors that determine the accuracy of the test results, so there will be no guarantee.
Q: What are the advantages of your instrument compared with those of Japan?
A: That’s a big problem. There are several different instruments in Japan. I don’t know which one to compare with. From the characteristics of our own instruments, our instruments have the highest cost performance, the best after-sales service, short delivery time, almost no consumable parts in the process of use, and low cost.
Q: What is the error range of the instrument?
A: There is no absolute error range, mainly for a certain order of magnitude of the sample set a relative error, for example: 1000PPM within the relative standard deviation rate is generally within 10%.
Q: What is the meaning of analytical accuracy 0.05% in technical indicators?
A: 0.05% refers to the general accuracy evaluation method of the instrument, which refers to the relative deviation of repeatability measurement of elements with a content of more than 96%.
Q: When the handheld spectrometer is testing an object, is there any deviation in the element content of the object if the Angle is changed?
A: There will be deviations, which is an important reason why the handheld X-ray fluorescence spectrometer test is coarser than the desktop X-ray fluorescence spectrometer.
Q: What is the difference between a spectrometer with a vacuum pump and one without one?
A: A spectrometer with a vacuum pump can measure light elements, while a spectrometer without a vacuum pump can measure major elements.
Q: How much does instrument radiation harm the human body?
A: The radiation produced by the instrument does not affect the human body (that is, the operator). Our instrument has been tested by relevant departments and can provide relevant safety test reports. At the same time, our instrument has a triple protection means to ensure that it does not cause any harm to personnel in use.
Q: Has it passed the national quality system certification?
A: Yes, we have passed the ISO9001:2008 national quality system certification.
Q: Can your instrument measure hexavalent chromium and bromine compounds?
A: We can’t measure hexavalent chromium and bromine compounds. We can only measure the total amount of chromium and bromine. This is determined by the principle of X-ray fluorescence. Any X-ray fluorescence spectrometer in the world can not do it, but we can provide qualitative test reagents for hexavalent chromium.
Q: Are your instrument standard samples authoritative?
A: Our instrument standard samples EC680K,EC681K are authoritative, can provide the relevant EU authority certification.
Q: Do you need any other auxiliary materials in the RoHS testing process?
A: No auxiliary materials or consumables are required.
Q: How is the instrument report generated?
A: The report of the instrument is divided into three report formats, namely Html file, Pdf file, Bmp file, and can also output reports in excel format
Q: Which industries is the handheld device suitable for?
A: It can be used in various industries that require portable testing, such as metallurgy, prospecting, metal recovery, etc. Its flexibility brings many benefits to customers.
Q: Why do count rates vary so much when RoHS is tested?
A: Because of different materials, X-ray fluorescence excitation efficiency of different materials is different, so the counting rate is not the same. At the same time, because the metal composition is more complex, the measured counting rate changes more, and its change does not affect the actual measurement results.
Q: How long is the sample test time?
A: The longer the measurement time, the higher the accuracy of the measurement, the general measurement time is related to the test count rate of the sample, the smaller the count rate, the longer the measurement time, the general test time 150S-300S, the count rate is smaller (less than 1000PCS), we recommend testing 300S is the best.
Q: Can liquids and inks be tested in your instrument?
A: Yes, our instrument can measure harmful metals in liquids, powders and solids (except gas), and the test method is very simple.
X-ray fluorescence spectrometer FAQ Series 1
Q: Why doesn’t the content result box pop up after the test?
A: Because the sample was named with special characters. Sample naming cannot use “V:,? Such special characters. This is limited by the Windows operating system.
Q: Why is a peak, half the color of mercury and half the color of lead?
A: Peak drift, this case should be re-initialized. It could also be interference from other elements, other than lead and mercury
Q: Why is it necessary to reduce the current to 400 mA when measuring stainless steel?
A: Because stainless steel is easily excited, the count rate is too high because the current is too large, which causes the spectrum accumulation phenomenon to be serious, and the high background makes the measurement error increase.
Q: Why can’t you measure the pin?
A: The pin can be tested. If it is tested for RoHS, it can be tested directly. If you test the coating, you need to use a small caliber collimator, and use the standard sample calibration, and at the same time to test the average of multiple points, so that you can meet the requirements of the test.
Q: Has your instrument passed the certification of the authority:
A: Yes, it has passed the inspection and certification of the national quality and technical supervision department. But the X-ray fluorescence spectrometer as RoHS testing, there is no national and international standards, so its authoritative certification in RoHS testing is no international standard, and, at present, there is no institution in the world can do RoHS measurement and testing certification.
Q: Will the results measured by your instruments be recognized by the state?
A: No, because the X-ray fluorescence spectrometer is a comparative analysis instrument, a crude measurement in RoHS testing, the measurement results are only used as a reference for the internal control of the enterprise, there is no authority, but in other industries have X-ray fluorescence detection standards, such as precious metal detection and archaeology and other fields, its measured results in some authoritative institutions can be recognized.
Q: What is the difference between your instrument testing and SGS testing? How much deviation: difference range?
A: X-ray fluorescence spectrometer is a physical method, but also a surface test, its RoHS is a rough measurement. Therefore, it has the advantages of non-destructive sample, fast, simple pretreatment and so on. SGS uses chemical methods, which are accurate measurements, but the early processing is complicated and the measurement time is long. For the detection of uniform materials, the deviation between the two is not large, and it is the same order of magnitude change, which can fully meet the purpose of rough measurement and rough analysis of RoHS.
Q: How about your company’s after-sales service?
A: Our company’s after-sales service is very good, has established a complete after-sales service system, for customer problems within 3 hours to respond,48 hours (according to the customer’s different regions) arrived, timely solve the problem, to give customers a satisfactory answer, special circumstances can change the instrument.
Q: The quality of your instruments is not good, according to the expert in making instruments. So I won’t buy your products at all. What should I do?
A: If it is convenient for you, you can tell me the name and phone number of that expert, and we can invite him to discuss this problem face to face. Our instrument has a lot of users, and we have been serving more than 400 customers of a big brand in the same industry in China. You can call the customers we serve, maybe the customer’s evaluation is closer to the truth.
X-ray fluorescence spectrometer FAQ Series 1
Q: Does your company have a special instrument for the detection of one element? For example: “lead”.
A: There are special instruments for detecting one element.
Q: Who are your major customers?
A: Samsung, Midea, Huawei and so on. I suggest you get the information from our sales department.
Q: The reliability of the instrument?
A: The reliability of the instrument is very high. That’s why it is supported by so many customers.
Q: Does the standard sample conform to EU standards?
A: Our standard samples are fully in line with the European Union standards, approved and approved by the European Union authority. We can provide the relevant certificates and materials.
Q: Explanation and explanation of some technical indicators of the instrument. For example: 1PPM — 99.99%.
A: 1PPM-99.99% refers to a measurement range, 99.99%6 is the constant 999900PPM. It refers to the minimum amount can be measured to 1PPM content (which has a direct relationship with the measured substance), the highest can be measured to 99… 99% of the content.
Q: What about the testing principles and methods of similar competitive products? What are their advantages over us?
A: The testing principle and method of similar competitive products are roughly the same. Our advantage lies in the very high cost performance of the equipment, as well as good after-sales service and customer reputation.
Q: Why do we use a collimator with a small diameter when measuring small samples?
A: When measuring small samples, it is possible to use a large collimator and its measurement results are more accurate. The use of small collimator is mostly to measure the content of a certain point of the sample, because the small diameter collimator can put the light more concentrated on the sample, can reduce the interference of the sample material next to the actual measurement of the sample, such as: measuring the solder joint on the PCB board, can reduce the influence of “green oil”.
Q: Why do similar elements interfere with each other?
A:Because the electrons of similar elements are excited with similar energies, their spectra will be superimposed on each other, and they will elevate the background of each other, so it is easy to interfere with each other.
Q: Why can’t cadmium be detected in tin?
A: Cadmium in tin can be measured, but because many tin standard samples do not have the standard content value of cadmium, so there is no measurement result at present. At the same time, because the tin peak interferes with the cadmium peak, the test result is greatly biased.
Q: What about the precision and accuracy of your instruments?
A: The accuracy of the instrument is different due to the difference in the sample and content. Taking plastic as an example, the deviation of the content of 1000PPM is about 50PPM; In the content of about 100PPM, the deviation taste is about 10PPM, the absolute deviation value of high content is large, but the relative deviation value is small; The absolute deviation of low content is small, and the relative deviation is large.
Q: Why can’t you measure the amounts of chromium VI, PBB and PBDE?
A: Because of the principle of X-ray fluorescence, X-ray fluorescence spectrometer can only measure the total content of elements in the substance, and its combining state can not be determined, so the content of hexavalent, PBB and PBDE can not be detected, but we can provide qualitative reagents for measuring hexavalent.
X-ray fluorescence spectrometer FAQ Series 1
Q: What requirements does the instrument have for the surrounding environment?
A: The ambient temperature is 15~30℃, it is best to have air conditioning, relative humidity s70%, the lower the better.
Q: For the same sample, why are your measurement results different from those of SGS?
A: The X-ray fluorescence test equipment belongs to surface test, and the surface penetration is different for different substances. SGS is a wet chemical analysis method, which is to dissolve the sample evenly and then measure. If the sample is homogeneous material, the white difference between the two is on the same order of magnitude. If the surface is coated with coating layer, then the measurement deviation between the two will be relatively large in the mixed test, and it should be measured separately.
Q: The count rate is low, why?
A: It has to do with the material, size and thickness of the sample. At the same time, there are contaminants in the detector or X-ray tube, which also have a great impact on the detector count rate.
Q: The curve has a negative slope. Why?
A: Possible causes: 1. The spectrum test of the standard sample is incorrect, or the content value is inputted incorrectly; 2. Artificial addition of non-standard sample points, that is, data error.
Q; Why the test, there is a high peak spectrum, but the result is ND?
A: This situation is due to the interference of other elements, whose interference has been corrected in the software, so the result is ND. Another case is that the calibration data in the software has errors or has been changed.
Q: Why is the spectrum of the samples tested before different?
A: 1. The curve data was changed; 2. The sample spectrum was inadvertently overwritten by another sample during testing.
Q: Why does the lead in the measurement of tin drift?
A: It is not a drift phenomenon, but an illusion of the spectrogram. That is because the element bismuth is mixed with tin, and the illusion is caused by the bismuth near the lead peak.
Q: Why does the RoHS software crash when the motor is moving during the test? Turn it off, turn it back on, and then work again?
A: This kind of phenomenon is very rare. It is most likely caused by the failure of Windows operating system when the test software crashes
Q: Why is the data lost after the RoHS test?
A: 1. The save path is incorrect 2. Abnormal shutdown 3. Special characters are used
Q: Why are the results of RoHS metal measurement very different from SGS?
A: 1. Spectrum classification is surface physical classification, SGS uses chemical dissolution sample classification method, the difference between the two measurement methods. 2. The error of the metal standard sample itself. 3. There is coating on the surface of the metal sample
Q: Why should the storage path of the spectrum be the same as the installation path of the software?
A: It is required when the software is designed.
Q: Why do I need to initialize the software every time I boot it up?
A: This can ensure that the hardware condition of the instrument is in the best condition.
Q: Why do I need to preheat the machine for 30 minutes? Is 10 minutes OK?
A: The instrument is fully preheated to ensure the accuracy of the test data. We recommend preheating for 30 minutes. At the same time, there is a quick preheating method, with our engineers training, can reach 10 minutes of preheating.
Q: Why is it that when using plastic plastic products to measure copper foil of PCB board, there is obviously a bromine peak, but Br will report ND?A: Because the peak of bromine is the green oil part, the software automatically filters out the influence elements of the non-metallic part.
Q: Why should Sn curve be selected for tin-coated metal materials instead of the fourth working curve?
A: Because X-ray fluorescence is a surface grading test instrument, if the metal surface is coated with tin, then the final result of the measurement, tin composition accounts for the main, if the type of metal is used to test, the result error will be large.
Q: Why is there a large deviation between the measured data and the standard content when measuring the standard sample EC680K?
A: Compared with EC681K, the content of EC680K is relatively low. For low content samples, the test data is greatly affected by the background, so the relative deviation of the test is large, but the absolute deviation is smaller. If the EC681K test deviation is not large, then there is no problem for RoHS test determination.
Q: Why is the deviation of the results of two consecutive measurements relatively large under the same conditions? Which result should we choose?
A: The deviation of the results of the general instrument under the same conditions is large, there are two reasons: 1. Caused by the unevenness of the sample. 2. The comparison data inside the instrument has been artificially altered. 3. The instrument received external interference, such as: interference from the power supply. 4. Hardware failure of the instrument. In the above faults, non-hardware faults, just re-proofread
.X-ray fluorescence spectrometer FAQ Series 1
Q: When the instrument is initialized, the tube flow, tube pressure, counting rate are normal, and the travel time is normal, but there is no silver peak. Why?
A: 1. Check whether the signal cable of the instrument is properly connected. 2. The interface of the instrument is interfered with, turn off the instrument and the computer for 5 minutes, then restart it and try again.
Q: When the high voltage failure, what phenomenon, why there will be high voltage power failure?
A: There is no counting rate, or the counting rate is very small, and there is an abnormal squeaking sound. There are many reasons for high voltage power supply failure. For example, the power supply voltage is unstable for a long time, the working environment is humid or the area with large pollution for a long time, etc., will cause the high voltage power supply failure.
Q: What is the phenomenon when the X-ray tube is broken?
A: No map, small or no counting rate, oil leakage and other phenomena.
Q: Why did the high voltage indicator light (yellow light) go off immediately when the test was stopped?
A: This is the hardware setting of the instrument. When the high voltage indicator goes off, it means that the high voltage stops the output of the high voltage. Of course, no X-ray is generated, which is a kind of safety protection for the human body.
Q: Why is there no time, no counting rate, and no spectrum during the test?
A: 1. Check whether the data connection cable is connected properly; 2. The microswitch on the cover of the instrument may not be in good contact with the same machine.
Q: What are the requirements for power supply?
A: The instrument needs to be equipped with a high-precision AC purification and stable power supply, because voltage fluctuations and interference will affect the test results. The operating voltage of the instrument is required to be within the range of 220±5V, and it is best to be equipped with an “online UPS” power supply.
Q: The peak spectrum in the test is very messy. Why?
A: Both the interface circuit and the amplifier circuit can cause such a failure, and software damage can also cause such a problem.
Q: The high voltage indicator is off or blinking when the light is on. Why?
A: High voltage discharge, or optical tube discharge, that is, the phenomenon of high voltage power supply discharge caused by various reasons can produce this phenomenon.
Q: The initialization is not in place, why?
A: 1. The magnification is set incorrectly in the software. 2. The DA chip of the LPU is damaged due to power supply. 3. The software is damaged and needs to be reinstalled.
Q: The test has no spectrum, why?
A: Is the microswitch in good contact with the upper cover of the instrument? At the same time, the optical tube, high voltage, interface board and detector are broken may cause no spectrum. If the software does not travel, it is likely that the instrument is not connected to the computer, or the instrument interface board is damaged.
Q: Why does the instrument squeak?
A: There is a possibility of discharge from the high voltage power supply or X-ray tube in the instrument.
Q: Why is the initial use of RoHS very long, peak channel does not reach 1105 stop, indicating the end of the initialization?
A: Generally, the detector magnification is not set well, or the initial parameters of the interface board are not corrected well.
Q: Why do you use different collimators for RoHS testing? What are the differences?
A: Different types of test samples require different test conditions, which have been set in our software, users do not need to set. Each kind of collimator corresponds to the test sample type is different. The main difference is the difference in the filter.
X-ray fluorescence spectrometer FAQ Series 1
Q: How to judge the filament, interface board is good or bad?
A: The high voltage end of the filament line unplug, use a multimeter to measure the resistance at both ends, resistance value is infinite, indicating that the filament has been broken; If the software shows “device end connection “, then the USB connection line is connected, generally the interface board fault.