X-ray energy dispersive fluorescence spectrometer XRF test using steps
Preparation of the Spectrometer
a) Energize the instrument, heat the equipment according to the working guide of the instrument, and stabilize the instrument according to the manufacturer’s instructions.
b) Ensure the stability of the test and make the detector stable according to the manufacturer’s instructions
2 Calibrate
a) Select the reference sample as described in section 7 according to the instructions in the instrument user’s manual. The concentration of elements in the sample must vary. If the calibration covers a lot of elements and the concentration range spans a large span, many calibration samples will be required. The number of calibration samples is reduced for the following reasons:
1. Calibration by basic parameter method (fewer elements than standard matter)
1. To calibrate (with a standard of a similar element) by basic parameters
1. Calibrate by basic parameter analysis plus experience
b) Analytical method calibration takes into account spectral interference, matrix effects and other effects, which all affect the determination of fluorescence scattering intensity in the spectrum. A list of these effects can be found in the annex to this chapter.
c) In order to ensure adequate analytical performance for each measurement element, the instrument must be placed under optimal measurement conditions by selecting the appropriate excitation parameters. These conditions are specific to the instrument. Typically, this information can be found in the analyst’s instruction manual.
d) As a general guideline, it is recommended that users of the method be aware that interference with the spectrum of the mutual elements and changes in the matrix between samples can fully affect the accuracy, precision, and minimum detection limits of the results of each analyte. For example, the detection limit for Cd in pure polyethylene can reach 15mg/Kg, but is not reached in the presence of 10% bromine compounds and/or 2% antimony
X-ray energy dispersive fluorescence spectrometer XRF test using steps
3 Detector Performance check
Whether or not the device meets the required performance standards, it needs to be determined by measuring certified reference materials or associated reference samples. The concentration of the element contained in the sample must not be greater than 3-5 times the screening limit for the test element (see Table 3). The result obtained from the reference sample must be within the margin of error allowed by the sample test. When equipment is used to analyze unknown materials, the manufacturer should provide a standard operating procedure (SOP) and an appropriate reference sample. The recommended method should ensure that the operator can obtain high quality analysis results.
4 Test the placement of the sample
a) If the sample to be tested can be placed in the sample chamber inside the benchtop X-ray fluorescence spectrometer, and the sample to be tested is in the appropriate measurement position, the corresponding test can be performed. If the sample to be tested does not fit in the sample chamber, it must be cut to size for measurement. In order for the test result to be valid, the instrument’s minimum requirements for sample thickness and quality must be met, as very small or thin samples may not meet this condition. In this case, such small samples (such as screws) should be placed in a sample cup and analyzed. Similarly, similarly thin samples should be stacked together so that the minimum sample thickness limit can be reached for effective analysis. A general rule is that all samples must fully cover the spectrometer’s measuring window and should be at least 5mm thick for polymers and light alloys such as Al, Mg and Ti, 15mm for liquids, and approximately 1mm for other alloys. However, the sample sizes required vary from device to device, and the spectrometer operator should consult the equipment manual or prepare the sample according to the minimum size/mass/thickness required by the manufacturer.
b) If testing on a portable XRF spectrometer, it must be noted that the measuring window of the spectrometer should be opposite and in direct contact with the test sample. Small or very thin samples should be operated as described in a) above. The analysis is then performed using an attachment (if available) that allows the operator to measure the sample in the sample cup. The regulations regarding the size/mass/thickness of the minimum sample also apply to portable analysis.
c) If the sample is liquid, powder, or spherical, or just a very small sample, it needs to be measured in a sample cup with a non-reusable film. When handling this window film, be careful not to touch its surface so as not to contaminate it.
X-ray energy dispersive fluorescence spectrometer XRF test using steps
5 Screening Tests
a) Test at the prescribed time.
b) Display and record the results of the analysis.
6 Analysis of results
The various types of XRF photometers are used separately, and the analysis method allows the user to classify the sample into one of the following three categories:
a) “Acceptable “- The sample has been tested as acceptable if the results of quantitative analysis for all elements are below the lower limits listed in Table 2.
b) “Unqualified” – If all of the elements are above the upper limit listed in Table 2, the sample is tested as unqualified.
c) “Undecided” — If the results of quantitative analysis for any of the elements Hg, Pb or Cd are in the middle region or if the results for the elements Br and Cr are above the high limits listed in Table 2, the results of the analysis are not decided. Progress must be madeFurther research. This test is “undecided”