GA-B116 Fully Automatic Radioactive Water Sample Evaporation, Concentration, and Acid Removal Apparatus
This radioactive water sample concentrator and acid removal instrument utilizes far-infrared radiation heating and PID temperature control.
- Opis
- Povpraševanje
Opis
Applicable Standards:
ISO 9696:2007 Water Quality – Determination of Total Alpha Activity in Salt-Free Water – Thick Source Method
ISO 9697:2008 Water Quality – Determination of Total Beta Activity in Salt-Free Water – Thick Source Method
EJ/T 1075-1998 Determination of Total Alpha Radioactivity in Water – Thick Source Method
EJ/T 900-1994 Determination of Total Beta Radioactivity in Water – Evaporation Method
DZ/T 0064.76-1993 Groundwater Quality Test Method – Radiochemical Determination of Total Alpha and Beta Radioactivity
HJ 898-2017 Water Quality – Determination of Total Alpha Radioactivity – Thick Source Method
HJ 899-2017 Water Quality – Determination of Total Beta Radioactivity – Thick Source Method
GB 8537-2008 Test Methods for Drinking Natural Mineral Water
GB/T 15265-94 Ambient Air – Determination of Dust – Gravimetric Method
GB/T 5750.13-2006 – Standard Test Method for Drinking Water – Radioactivity Index
GB/T 5750.4-2006 – 8.1 – Water Quality – Determination of Total Dissolved Solids – Standard Test Method for Drinking Water
Opis izdelka:
This radioactive water sample concentrator and acid removal instrument utilizes far-infrared radiation heating and PID temperature control. During the experiment, a peristaltic pump and a high-precision weighing module are used to achieve precise quantitative concentration. It can simultaneously concentrate 1-4 vzorcev, enabling unattended operation and improving experimental efficiency. Suitable for the water quality and tap water industries, it provides fully automated pre-treatment for water sample evaporation, concentration, and acid removal during the testing of total alpha, beta, and other radioactive water samples; automated evaporation and concentration of ambient air dust samples; and evaporation and concentration of total dissolved solids (TDS).
Tehnični parametri:
Model: GA-B116
Concentration Temperature: Room Temperature – 600°C
Acid Addition Temperature: Room Temperature – 600°C
Concentration Sample Volume: 1-9999ml (Single Channel)
Concentration Endpoint (ml): Individually adjustable based on experimental requirements
Single Liquid Dosage (ml): Adjustable based on experimental evaporating dish size
Cleaning Time (s): Adjustable based on requirements
Concentration Units: 1-4, individually controlled, non-interference units
Rated Power: 2000W
Evaporation Vessel: Compatible with 125ml/200ml/300ml porcelain evaporating dishes
Intelligent Alarm, Dual Voice and Text Alarm
National Standard Method, No Sample Transfer Required, Minimizing Transfer Sample Loss
Four independent channels, one-button start, automatic concentration and quantitative measurement
Single-port, single-control, imported far-infrared radiation heating, PID temperature control
High-precision sensor quantitative measurement, intelligent water sample addition, individually adjustable sample volume, automatic heating stop upon reaching the set value
Automatic cleaning, four-channel automatic cleaning, individually adjustable cleaning times
The entire unit is corrosion-resistant and heat-resistant, ensuring long-term stable operation.
It can be upgraded to automatic acid addition, enabling unattended operation and improving experimental efficiency.
Unit Dimensions: 600mm x 395mm x 345mm