Design Standard:
| GB11158 | High Temperature Test Chamber Technical Condition |
| GB10589-89 | Low Temperature Test Chamber Technical Condition |
| GB10592-89 | High-low Temperature Test Chamber Technical Condition |
| GB/T10586-89 | Damp Heat Test Chamber Technical Condition |
| GB/T2423.1-2001 | Low Temperature Test Chamber Test Methods |
| GB/T2423.2-2001 | High Temperature Test Chamber Test Methods |
| GB/T2423.3-93 | Damp Heat Test Chamber Test Methods |
| GB/T2423.4-93 | Alternating Damp-Heat Test Methods |
| GB/T2423.22-2001 | Temperature Changing Test Methods |
| IEC60068-2-1.1990 | Low Temperature Test Chamber Test Methods |
| IEC60068-2-2.1974 | High Temperature Test Chamber Test Methods |
| GJB150.3 | High Temperature Test Methods |
| GJB150.4 | Low Temperature Test Methods |
| GJB150.9 | Damp Heat Test Methods |
GA-2001-D Programmable Temperature Humidity Chamber225L
Uses:
This equipment can simulate different environment condition. It’s appropriate for testing material performance, such as resist heat, resist dry,resist humidity and resist cold. That can define performance of material.
GA-2001-D Programmable Temperature Humidity Chamber225L
Parameters
| 1. Item
2. Model |
Constant Temperature and humidity test chamber
225L |
| 3. Function | It simulates the products in the chamber where the climate condition of different temperature and humidity, such as high-low operation and storage, temperature cycling, high-low temperature and high-low humidity, moisture condensation etc. to check its adaptability and changing performance.
It’s required to conform to international regulation (IEC, JIS, GB, MIL…) to reach the consistency (including test schedule, condition, method) under the international standard. |
| 4. Sample application | The chamber applies to:
compatibility test for store, transportation and application of electric, electronic products, unit components, parts and other materials under high or low temperature. The following samples are prohibited to be tested and stored: – Inflammable , explosive and other volatile substances; – Corrosive substance; – Biont; – Strong electromagnetic emission source. |
| 5. Volume, size & weight | |
| 5.1. Volume
5.2. Internal size 5.3. Exterior size 5.4. Weight |
225L
W600*H700*D500mm W1150 x H1800 x D1080mm 300KG |
| 6. Performance | |
| 6.1. Test condition | Temperature:+25℃, Humidity: ≤85%, no samples in the chamber |
| 6.2. Test method | GB/T 5170.2-1996 Temperature test chamber
GB/T5170.5-1996 Temperature and humidity test chamber (only temperature and humidity) |
| 6.3. Temperature range | -40℃→+150℃ (adjustable) |
| 6.4. Temperature fluctuation | ±0.5℃ |
| 6.5. Temperature accuracy distribution | ±2.0℃ |
| 6.6. Humidity range | 20%~98%R.H (temperature at 20℃-85℃) |
| 6.7. Humid. fluctuation | ±2.5%R.H. |
| 6.8. Humidity accuracy distribution | ±3.0%. |
| 6.9. Heat-up time | -40℃ → +150℃ Heat up for 2-4.0℃/min |
| 6.10. Cooling time | +150℃→ -40℃ Cooling for 0.7-1.0℃/min |
| 6.11. Load condition | Testing space is not more than 3/4 of the total space, no heating |
| 6.12. Sample holder | Stainless steel punching stencil |
| 6.13. Relative humidity and temperature control range (refer to fig.) | ![]() |
| 6.14. Meet the technical standards
|
GB/T 2423.1-2001 Test A: Low temperature test method
GB/T 2423.2-2001 Test B: High temperature test method GJB 150.3-1986 High temperature test GJB 150.4-1986 Low temperature test IEC68-2-1 Test A: Cold IEC68-2-2 Test B: Dry heat GB11158 “Technical Conditions for High Temperature Test Chamber ” GB10586-1 “Technical conditions for Damp heat test chamber ” GB/T 2423.2 “Basic environmental testing procedures for electric and electronic products B: High temperature test method”. GB/T 2423.3 “Basic environmental testing procedures for electric and electronic products Ca: Constant damp heat testing method” etc. |
| GA-2001-D Programmable Temperature Humidity Chamber225L
7. Structure characteristics. |
|
| 7.1. Insulated enclosure structure
|
External material: SUS304 or stoving varnish
Inwall material: SUS304 Insulation material of box body: flame retardant fire rigid polyurethane foam Door thermal insulation material: flame retardant fire rigid polyurethane foam |
| 7.2. Air conditioning channel | Centrifugal fan:
Heater, evaporimeter, outfall, temperature sensor |
| 7.3. Test chamber standard
|
Window 330x450x40mm 3 vacuum layers
Anti-explosion handle Door hinge: SUS #304 Lamps in the chamber: 11W Access port: φ50mm or φ100mm 1pc or more. |
| 7.4. Door
|
Single door, open from left, handle is at right; With viewing window: W300xH450mm, lamp window/condensation prevention thermal device |
| 7.5. Controller | Control screen, overtemperature protection setting |
| 7.6. machine room | Refrigeration units , water pans , outfall |
| 7.7. control cabinet for power distribution | Switchboard fan, ventilator |
| 7.8. Heater | Heating system: cycling heating from finned-tube Stainless steel electric heater |
| 7.9. Humidifier | Oven humidification system, cycling winding system |
| 7.10. Power line hole and outfall position | located at the back of chamber |