ESD30K Electrostatic Simulation Generator
- Description
- Inquiry
Description
ESD30K Electrostatic Simulation Generator
Product Overviews
Electrostatic discharge (ESD) is a common natural phenomenon. The effects of ESD on electrical and electronic equipments, devices or systems are everywhere. It is a very high degree of harmful electromagnetic energy. Only to improve the level of anti-static interference can ensure the safe use of electronic products. ESD generator is used to evaluate the electrostatic immunity performance of electrical components and electronic equipments by simulating real electrostatic discharge as defined in the standard IEC/ EN 61000-4-2. Kindly find below features & technical data of our model ESD 30K.
ESD30K Electrostatic Simulation Generator
Features
- User friendly 7” Touch Panel Display
- Support multi-language & facilitate users
- Built-in environment self-test program
- Programmable operation
- Pre-programmed IEC/ EN 61000-4-2 test settings
- Very convenient to replace the discharge modules to meet different standards of the test requirements
- RS232 / USB Port, PC control operation & printable test report
ESD30K Electrostatic Simulation Generator
Technical data
| Model | ESD30K |
| Standard | IEC61000-4-2,GB/T 17626.2 |
| Operation | 7” Touch Panel Display |
| Output voltage | 0.2~±30kV |
| Polarity | Positive/ negative/ alternating |
| Energy storage capacity | 150pF |
| Discharge resistor | 330Ω |
| Discharge current rise time | 0.6~1ns |
| The first peak current (±15%) | 7.5A at 2kV/ 15.0A at 4kV/ 22.5A at 6kV/ 30.0A at 8kV |
| Current at 30ns (±30%) | 4.0A at 2kV/ 8.0A at 4kV/ 12.0A at 6kV/ 16.0A at 8kV |
| Current at 60ns (±30%) | 2.0A at 2kV/ 4.0A at 4kV/ 6.0A at 6kV/ 8.0A at 8kV |
| Holding Time (air discharge) | ≥5s |
| Discharge Modes | Contact Discharge & Air Discharge |
| Operating Modes | Single discharge & repetitive discharges |
| Discharge Trigger | Manual, automatic or remote |
| Working Mode | IEC level mode/ user mode/ program mode/ voltage rising patterns |
| Discharge Interval | 0.05~9.99s |
| Discharge Count | 1~9999 |
| Power Supply | AC220V ± 10%,50/60Hz |
| Ambient Temperature | 15℃~35℃ |
Environment
- Indoor use
- Altitude not exceeding 1000metres
- Ambient temperature 15℃~35℃
- Relative humidity no more than 85%
- No conductive dust, no fire or explosion hazard, no corroding metal or insulating gas, sine wave voltage waveform, waveform distortion rate≤5%
- Earthing resistance not more than 0.5Ω



















