X-ray fluorescence Spectrometer FAQ Series 2

X-ray fluorescence Spectrometer FAQ Series 2

Customer frequently asked questions about XRF spectrometer:

1. How to maintain the Energy dispersion XRF ? What should be done with X-ray tubes when sample analysis is not carried out?

Mainly spectrometer room constant temperature and humidity (o 22 stopnie, humidity below 60%), clean environment, stable power supply, avoid vibration. When the sample analysis is not carried out, the X-ray tube current and voltage should be reduced to the minimum, and the instrument should be kept at a constant temperature

5. If there is no analysis for 1-2 godziny, should the X-ray tube be turned off?

NIE. The X-ray tube can not be casually closed, for a long time you need to re-age, generally you can reduce the power down when you do not use: voltage 20KV, current 10mA.

X-ray fluorescence Spectrometer FAQ Series 2

7. The X-ray tube has two kinds of side window & end window. Anoda szyby bocznej jest uziemiona, a żarnik jest połączony z ujemnym wysokim napięciem. Okno końcowe jest uziemione przez żarnik. Woda używana do chłodzenia tarczy końcowej rury okiennej musi być czystą wodą o dużym oporze i używać urządzenia z obiegiem wody wyposażonego w żywicę jonowymienną, natomiast typ okna bocznego może być używany do wody pitnej.

Czy mogę zapytać, dlaczego do okna końcowego należy zastosować czystą wodę o wysokim oporze? Jakie jest zastosowanie urządzenia z obiegiem wody zawierającego żywicę jonowymienną?

(1) Ponad 99% energii wytwarzanej przez świetlówkę jest zamieniana na energię cieplną, duża ilość energii cieplnej musi zostać odebrana wraz z krążącą wodą, nie można używać wody ogólnej, musi to być woda dejonizowana o wysokiej czystości, lub przewodzący, świetlówka może zostać zezłomowana, żywica jonowymienna jest dejonizowana, Na przykład, właśnie zmieniliśmy wodę obiegową, gdy zobaczyliśmy, że przewodność jest bardzo wysoka, biegnąc przez chwilę w dół, taka jest rola żywicy jonowymiennej.

(2) Gorące elektrony wytwarzane przez katodę (włókno) lampy rentgenowskiej są przyspieszane do anody (cel) pod działaniem pola elektrycznego wysokiego napięcia Więcej niż 99% energii zamienia się w energię cieplną, duża część energii cieplnej musi zostać odebrana wraz z krążącą wodą, nie można używać wody ogólnej, musi to być woda dejonizowana o wysokiej czystości, lub przewodzący, świetlówka może zostać zezłomowana, żywica jonowymienna jest dejonizowana, Na przykład, właśnie zmieniliśmy wodę obiegową, gdy zobaczyliśmy, że przewodność jest bardzo wysoka, bieganie przez jakiś czas zeszło, to jest rola żywicy jonowymiennej. Woda dejonizowana bezpośrednio chłodzi cel anodowy. Zewnętrzna rura chłodząca lampy rentgenowskiej jest wykonana z metalu, podczas gdy wewnętrzna rura chłodząca do izolacji elektrycznej ma długą zwiniętą rurkę nylonową (w celu zwiększenia odstępu izolacyjnego). Katoda (włókno) i tarcza anodowa są zamknięte w komorze wysokiej próżni (aby zapobiec utlenianiu i zmniejszyć tłumienie promieni rentgenowskich), a tarcza anodowa jest zamocowana za pomocą szklanej osłony, a olej izolacyjny jest izolowany (skorupa gniazda kabla wysokiego napięcia i spiralna rurka nylonowa są zanurzone w oleju izolacyjnym wysokiego napięcia). Dlatego, obecność jonów metali w wodzie chłodzącej nie jest konieczna, aby nie spowodować przewodzenia prądu lampy rentgenowskiej do masy, wyładowania i awarii. Wskaźnik izolacji wody dejonizowanej wynosi 2us/cm.

X-ray fluorescence Spectrometer FAQ Series 2

11. Zrób złoto, oznaczanie platyny musi dotyczyć próbki standardowej? Złoto to biżuteria do rąk zwykłych użytkowników.

(1) Zrób złoto, oznaczenie platyny musi dotyczyć próbki standardowej, chyba że wyniki oznaczania się nie liczą

(2) Naszyjnik należy najpierw zdezynfekować

12. Use XRF to measure the lead in the magnetic core material, the requirement is below 100PPM, can it be measured?

100PPM below the lead content can be measured, of course, can not be too small, such as a few PPM is more difficult to test, want to have an accurate quantitative must buy a standard sample, the standard sample must have lead.

18· Energy type and wavelength type compared with what are the advantages and disadvantages?

The energy analysis area is relatively small, the analysis accuracy is also almost, and the speed may be slower, but the operation is simple

21.X fluorescence spectrum thermoelectric instrument, the role of the detection mouth put a copper sheet?

(1) Some handheld devices contain a corrected sample in the head, and can also protect the X-ray emission source when not in use.

(2) Rolą bloku miedzi jest dokonanie korekty energetycznej, a drugim jest blokowanie okna, aby chronić inne próbki, gdy nie zostaną wykryte.

X-ray fluorescence Spectrometer FAQ Series 2

22. Dlaczego fluorescencja rentgenowska określa zawartość Sb w próbce tabletki, gdy ma wpływ grubość próbki??

(1) Element (lub matryca) o niższej liczbie atomowej ma niższy współczynnik absorpcji dla linii widmowej o wyższej energii, więc nieskończona grubość jest trochę większa

(2) Energia linii K Sb jest wyższa i może przeniknąć przez grubszą próbkę (względem charakterystycznych promieni innych pierwiastków w próbce), więc grubość nasycenia jest grubsza niż w przypadku pozostałych elementów.

225. Podstawowa metoda kalibracji spektrometru fluorescencyjnego?

Do korekty użyj standardowej próbki

27. Ile jest rodzajów spektrometrów fluorescencji X? What is the difference between energy dispersion and wavelength dispersion?

(1) The characteristics of each element X-ray fluorescence has different wavelength and energy, wavelength dispersion and energy dispersion is the use of its wavelength and energy difference to detect.Wavelength dispersion uses crystal splitting to separate characteristic X-rays of different wavelengths, and makes qualitative analysis of elements according to Bragg formula and Moseley’s law; According to the higher the element content, the stronger the intensity of the ray, through a certain method to calibrate and correct, and finally quantitative analysis.Energy dispersion does not use crystal splitting, directly use a semiconductor detector to detect the characteristic X-ray energy of the sample for qualitative and quantitative analysis

(2) There are energy, wavelength, total reflection, synchrotron radiation, X-ray microfluorescence, proton excited X-ray fluorescence.

28. How does X-ray count gas?

The X-ray ionizes the gas. Under the action of electric field, the ionized electrons and positive ions move towards the poles respectively. In the process of electron movement towards the anode, they are gradually accelerated to obtain higher kinetic energy. When these electrons collide with the gas molecules, they will cause further ionization, resulting in a large number of electrons surging to the anode, producing an avalanche effect, which plays an amplified role. Each photon actually produces an avalanche, and the number of photons can be counted by using a counter to detect the resulting electrical pulse

X-ray fluorescence Spectrometer FAQ Series 2

29. Has any friend ever used EDX to detect Si (about 0.05~0.1%) and Fe(about 0.05~0.5%) in pure aluminum? What is the method?

(1) Fe should be easy to measure

Si in pure AI is difficult to measure for the following reasons:

The difference in the content of Al and Si is nearly a thousand times, and the characteristic ray energies of Al and Si are too close together, and the background (which is a random quantity) generated by the high content of Al at the position of Si is enough to make the intensity of Si change dramatically

(2) Light elements in the air spectral lines will be strongly absorbed, resulting in spectral peaks do not come out or the intensity is very low. In the vacuum state will be better.

(3) If Ka is used, Al is 1.487 and Si is 1.740, which will easily cause interference.

30. What is inside the molecular pump and why is it more expensive than the vacuum pump?

Mówisz o pompie turbomolekularnej? B: Pompa próżniowa składa się z pompy mechanicznej, pompy turbomolekularnej i pompy dyfuzyjnej, pompa mechaniczna pompuje głównie niską próżnię, pompa turbomolekularna pompuje wysoką próżnię.

31. W przypadku stosowania spektrometru fluorescencyjnego XRF do pomiaru Pb istnieje linia analizy PbLb1. Czy istnieje linia analizy PbLb2?

(1) La i Lb1 Pb to dwie najbardziej wrażliwe linie. Czułość pozostałych linii jest bardzo niska, i nie jest wygodny w rutynowym użyciu.

(2) Istnieje linia analizy PbLb2, ale jego siła jest zbyt słaba, to znaczy, czułość względna jest bardzo niska, w normalnych warunkach wybierana jest linia Pb Lb1.

32.Jak połączyć wartość szczytową linii widmowej, aby zobaczyć wyniki testu EDX? Jeśli zmierzona zawartość Pb jest wysoka, ale odpowiadający mu pik Pb jest mały, what is the explanation?

(1) I don’t know when you say the content is high, how high is it?My humble opinion: Przede wszystkim, we must see if there is any interference from other elements, such As AS interference is relatively strong.Second, look at the peak value of the spectral line, not the height.

(2) As far as I know, the content of elements is not as high as the peak, and the excitation efficiency of the elements in the sample is related. Na przykład, if your DT is 20%, the peak of Pb is very high at that time, but the content is not necessarily high because other elements are not excited. So be sure to look at DT between 40% I 50% is the best detection results

X-ray fluorescence Spectrometer FAQ Series 2

33. How long is the EDXRF correction cycle?

(1) PHA(spectral drift) correction, α(instrument drift) correction need to be done every day.

(2) The above correction must be done before quantitative analysis.

(3) Also check the calibration! There are sensitivity library correction!

(4) At the beginning of your operation, you can congenital days to do PHA correction, and then stable week/time!

(5) Drift correction I generally do not do, one instrument is more stable, two to feel a little trouble. It is recommended to make a monitoring sample before measuring the sample.

34. Is a standard sample of all components required for quantitative testing? What I want to test is the glass sample. What is the reason why the components cannot be measured?

Is it because of organic matter?General EDXRF can test the elements between NA-U, and the elements in front of Na are powerless, such as H, O and other elements, so if it is to test organic matter, it can not be tested.

35. Dlaczego problem z przedwzmacniaczem wpływa na inne elementy?

Problem ze stałym kanałem będzie oczywiście dotyczył pozostałych elementów, ponieważ inne elementy mogą wykorzystać siłę problemu do obliczenia współczynnika korygującego, co jest ze sobą powiązane.

36. Jaka jest metoda krzywej standardowej dla ED-XRF? Jak mierzy się krzywą roboczą?

(1) Metoda krzywej roboczej polega na użyciu próbki standardowej w celu ustalenia standardowej krzywej korekcji, a następnie obliczyć wyniki pomiarów zgodnie z tą krzywą podczas testu.

(2) Przy znanym stężeniu kilku standardowych próbek do pomiaru wytrzymałości, uzyskać serię odpowiedniego sygnału siły, kolumna współrzędnych kolumny, siła i stężenie odpowiadające utworzeniu szeregu linii punktowych, to jest krzywa standardowa, przetestować próbkę, przyrząd sprawdza siłę sygnału, odpowiadający tej krzywej może znać odpowiednie stężenie.

(3) Metoda krzywej standardowej musi najpierw mieć pewną znaną zawartość próbki standardowej, ustawić warunki testu po wejściu na standardową zawartość próbki, a następnie przetestuj próbkę standardową, można uzyskać wartość siły sygnału, wartością siły jest współrzędna pozioma, zawartość to współrzędna pionowa, za pomocą której tworzy się krzywą, po instrumencie testowym Wartość siły zmierzona bezpośrednio przez instrument może zostać przeliczona na zawartość

(4) Metoda krzywej standardowej dla XRF polega na użyciu próbki standardowej o znanej zawartości, ustawić warunki analizy, wprowadź dokładną treść, użyj przyrządu do określenia wytrzymałości próbki, siłą jest współrzędna pionowa, zawartość to współrzędna pozioma, poprzez zastosowanie programowego doboru równań regresji, siłaKrzywa standardowa jest syntetyzowana z zawartością. Krzywa ta służy do obliczenia krzywej zawartości po przeanalizowaniu siły wzoru przez instrument.

(5) Metoda standardowej krzywej roboczej, zwana także metodą EC, to skrót od Equivalency of Concentration. Metoda standardowej krzywej pracy jest dokładniejsza niż metoda FP, ale ze względu na siłę tła samego instrumentu, próbka poniżej siły tła będzie ujemna. Wartość, dryft standardowy i wytrzymałościowy wpływają na dokładność krzywej, granica wykrywalności krzywej i górna granica krzywej. Aktualne standardowe krzywe robocze dla substancji niebezpiecznych zgodnie z dyrektywą RoHS to:

1, PVC/PE 2, stop miedzi 3, aluminum alloy 4, lead-free solder

X-ray fluorescence Spectrometer FAQ Series 2

37· When we use powder pressing, we use a general standard sample, to jest, to make the standard sample reach a certain particle size.Is there a standard sample of spectrum professional? Is it the same as what we use now?

(Mainly powder samples, here the spectrum also mainly refers to X-ray fluorescence)The so-called spectral standard sample refers to the standard sample suitable for spectral analysis, not such as the grade difference of reagentsThe process of chemical purity – analytical purity – superior purity – high purityGeneral spectral standard samples are not versatile because they are expensiveOr the shape of the spectrum standard sample is blocky, cutting or remaking the sample will cause damage to the original.Therefore, jest całkowicie odpowiedni dla krajowych próbek standardowych powszechnie stosowanych w laboratoriach

38. Jaka jest czystość oznaczonej próbki “widmowo czysta”?Czystość widmowa nie została wymieniona we wcześniejszym podręczniku klas odczynników chemicznych

W zasadzie pojawił się dopiero w ostatnich latach, to, co nazywa się czystym odczynnikiem spektroskopowym, mieści się w przedziale pomiędzy wysoką czystością a doskonałą jakością, ponieważ wysoka czystość jest droga, niektóre standardowe próbki nie mają tego argumentu najwyższej klasy. Nie może osiągnąć czystości o wysokiej czystości, ale widmo można wykorzystać, i jest opisywany jako czysty widmowo

39. Czy dokładny jest pomiar zawartości metali ciężkich w częściach metalowych metodą XRD??

(1) Jest OK. Pod warunkiem, że zawartość pierwiastków nie jest bardzo mała i pik widma nie jest zasłonięty przez tło, ogólny pomiar jest nadal OK. Jeśli źródło minerałów jest złożone, a metodą topienia szacuje się, że zawartość jest wysoka, wadą jest to, że topienie jest zasadniczo rozcieńczone, więc nachylenie krzywej nie jest wysokie.

(2) XRF może testować jedynie całkowitą Cr. Nie potrafi rozróżnić stanu wartościowości. XPS potrafi rozróżnić stany walencyjne. Czy głębokość testowa jest zbyt płytka dla ROHS

40. Jaki jest standardowy zakres temperatur i wilgotności w środowisku pracy? Jaka jest temperatura i wilgotność środowiska pracy XRF? B: Pierwszy, porozmawiajmy o naszych. Nasze środowisko pracy XRF ma temperaturę 15-30 stopni Celsjusza, a wilgotność osiąga szczyt powyżej 90%6 (w zeszłym roku o tej porze) ale ten rok jest lepszy i obecnie wynosi ogólnie około 70%6– 80%. Myślę, że wilgotność jest zbyt wysoka, raz wspomniany liderowi, ale wydaje się, że lider nie jest skłonny kupić osuszacza w celu zmniejszenia wilgotności. Wśród nich, gdy temperatura przekroczy 28 stopnie,XRF często generuje alarm temperatury w komorze próżniowej, więc ogólnie jest to kontrolowane poniżej 28. Jaki jest ogólny zakres Twojej temperatury i wilgotności

(1) Mój XRF jest umieszczony w wylocie klimatyzacji, temperatura jest ok 20, a wilgotność w pomieszczeniu utrzymuje się na poziomie 50-60

(2) Temperatura pokojowa naszego XRF jest kontrolowana na poziomie 18 ~ 28 stopni Celsjusza (najlepsza rekomendacja to 21 ~ 25 stopni Celsjusza), a wilgotność względna jest kontrolowana w temp 40% ~70%6. W rzeczywistości, temperatura jest względna, jeśli chodzi o działanie instrumentu, nie jest zbyt duża, przede wszystkim należy dobrze kontrolować wilgotność.

(3) Nasze wymagania dotyczące temperatury są kontrolowane pomiędzy 20-25 ° C, a wilgotność nie może przekraczać 70% najwyżej. The installation engineers have especially emphasized that failing to meet the requirements is likely to shorten the service life of the X-ray tube

(4) The indoor temperature is within 10 ° C do 35 ° C, and the humidity is 35% Do 80%

X-ray fluorescence Spectrometer FAQ Series 2

42. How to do Alpha correction?

There is an operation called standardization or alpha calibration, which is to measure the standardized sample used for its corresponding working curve before analyzing the sample to be tested (one or two standard samples must be selected as standardized samples or alpha correction samples when building the working curve). To correct for daily variations in X-ray intensity.

44. What do you mean by qualitative and quantitative analysis?

(1) The so-called qualitative analysis refers to the analysis to determine the nature of the substance, to jest, w celu określenia składu substancji przeznaczonej do analizy za pomocą różnych metod badawczych; Analiza ilościowa, odnosi się nie tylko do określenia składu substancji, ale także w celu określenia ilości lub jakości różnych komponentów, procent i inne istotne informacje.

(2) Analiza jakościowa: określić, jakie elementy, substancja zawiera jony lub grupy funkcyjne. Analiza ilościowa: określić zawartość odpowiednich składników w substancji. Słownik chemiczny, Wydawnictwo słownikowe w Szanghaju, Wrzesień 1989.

(3) Tak zwana analiza jakościowa ma na celu określenie, jaki element, analiza ilościowa polega na określeniu zawartości pierwiastka

47. Ze względu na potrzeby pracy, obecnie często spotykane żelazo powlekane (śruby, baterie), oraz zastosowanie analizy parametrycznej, rezultatem jest często bardzo mała zawartość żelaza, i cynk w powłoce, zawartość niklu jest bardzo duża. Jaką metodą można dokładniej zmierzyć pierwiastki zawarte w zakazie RoHS?

Myślę, że niski poziom żelaza jest normalny z dwóch powodów

:1) Substratem jest żelazo, padające promieniowanie rentgenowskie po absorpcji powłoki, dotarcie do warstwy żelaza zostało osłabione, więc wzbudzona fluorescencja żelaza jest mniejsza

.2) Zwłaszcza w przypadku filtra, intensywność fluorescencji rentgenowskiej Fe będzie niższa po prześwietleniu padającego promieniowania rentgenowskiego. W rzeczywistości, zawartość Fe jest niewielka, ponieważ rzeczywistą częścią podlegającą pomiarowi jest powłoka, zamiast żelaznej matrycy. Jeśli żelazną matrycę można zeszlifować, pozostawienie tylko powłoki na wymiar jest idealne, ale w praktyce nie da się tego zrobić. Metoda FP jest jak powiedzieć “półilościowe” metoda, Myślę, że nawet “półilościowe” nie ma o czym rozmawiać. Niezależnie od tego, czy jest to metoda FP, czy metoda linii pomiarowej, Nalegam, aby najpewniejszą oceną było najpierw ustalenie, czy w widmie znajduje się szczyt materii obiektowej, a następnie wyniki ilościowe. Weź śrubę, jeśli jest to produkt ocynkowany, jeśli powłoka zawiera Cr w kilkudziesięciu ppm (Śruby ocynkowane zazwyczaj zawierają Cr, ale nie mogę określić, czy tak jest 6 ceny lub 3 ceny), to uważam, że widmo musi być w stanie zobaczyć szczyt Cr. A jeśli są to śruby niklowane, w widmie nie będzie piku Cr. Załącznik stanowi przykład naszego testu, częścią jest także śruba, dawniej pokrywano cynkiem, teraz jest pokryty Ni, widać, że siła piku Fe w widmie jest również bardzo niska, i nie ma piku Cr (CrKa1 5,41KeV, CrKb1 5,95KeV).

48. Kiedy do wytwarzania metalu stosowana jest metoda FP, Wartość Cd jest bardzo wysoka, ale nie ma piku Cd w widmie, jak to wyjaśnić?Jakiego materiału bazowego używasz? Czy konfiguracja jest prawidłowa?

Ogólna metoda FP nie bada Cd i Hg, testuj tylko Pb i CrPrzed ustaleniem warunków metody FP, konieczne jest użycie “jakościowo ilościowe” aby zeskanować elementy składu materiału, jeśli zawiera składniki Sn, wówczas zmierzona wartość Cd jest często zbyt wysoka. Ponieważ linie widmowe Cd zostaną osuszone przez linie widmowe Sn. W metalu wystąpi również ingerencja ołowiu, sprawdź, czy próbka testowa ma wysoką zawartość ołowiu.

X-ray fluorescence Spectrometer FAQ Series 2

49. Czy XRF może wykrywać metale ciężkie w żywności?? Chcesz monitorować rynek, potrzeba szybkiego ustalenia, łącznie z obróbką wstępną, analizę najlepiej przeprowadzić w ciągu godziny!Nie musisz być zbyt dokładny, poziom ppm!Obiekt testowy na surowe mięso, ryba, kilka warzyw i owoców co, punkty Elementy analizy rtęci, Ołów, chrom, kadm i tak dalej!

Można również wykryć dziesiątki PPM więcej niż XRF, zawartość metali ciężkich w żywności jest dość mała, najniższe na świecie 0,5 ppm, Jak zmierzyć fluorescencję promieni rentgenowskich

50. Problem starzenia się lamp rentgenowskich, dlaczego ma się starzeć? Jakie są konsekwencje braku starzenia się? I jak się starzeć? Czy starzenie się wpływa na żywotność świetlówki??

(1) Po pewnym czasie bezczynności lampy rentgenowskiej, jego podciśnienie wewnętrzne zostanie zmniejszone, jeśli nie został przeszkolony i bezpośrednio podniesiony do wyższego kV, Wartość mA spowoduje wyładowanie pod wysokim napięciem, poważny, doprowadzi do uszkodzenia lampy rentgenowskiej! Dlatego, po pierwszym okresie bezczynności, be sure to slowly alternately increase the voltage and current (each interval of 5 minuty), and the new instrument basically has automatic training function, you can rest easy.

(2) Where there is damage to the X-ray tube is when it is turned on and off, the aging of the boot and the cooling of the shutdown will have an impact on the life of the X-ray tube.

51. Why does the X-ray tube have some upper irradiation and some lower irradiation? What is the difference in application?This is in the research instrument, instrument company gave me to share with you.

A Advantages and disadvantages of the downbeam spectrometer:Zalety: can analyze pure liquid samples; The lower surface of the sample is the analysis surface, and the bubbles generated after the liquid sample is heated have no effect on the analysis.Defects: Once the sample is broken during the analysis of powder tablet, the test room will be polluted and the X-ray tube will be damaged! It is difficult to deal with the debris falling on the window of the X-ray tube during daily analysis, which is not conducive to instrument maintenance;

B Advantages and disadvantages of the upper irradiation spectrometer:

Defects: can not analyze pure liquid samples; Only other methods can be used to analyze liquid samples;

Zalety: conducive to instrument maintenance, protect the X-ray tube is not damaged;

There is no difference in the analysis accuracy of the two types of instruments.

52. If the products of copper and iron matrix are tested by ROHS directive, what method is used?

(1) To use the method of analyzing heavy metal elements

(2) Copper alloy lead, kadm, mercury, Oznaczanie sześciowartościowego chromu i bromu można przeprowadzić poprzez trawienie kwasem, próbki wody za pomocą ICP,AAS lub UV do ustalenia, trawienie rtęcią i bromem w celu uszczelnienia trawienia, a następnie wybierz odpowiedni instrument, w zależności od siły ekonomicznej, aby kupić instrumenty Oprócz trzech rodzajów ICP dla sześciowartościowego chromu, wszystkie elementy, AAS dla ołowiu, kadm, UV wszystko można zrobić

(3) Produkty z osnową metaliczną, takie jak miedź, żelazo można bezpośrednio zbadać za pomocą XRF, jeśli występuje powłoka, można ją zeskrobać za pomocą analizy czystego kwasu rozpuszczonego w XRF w celu pomiaru obu wartości, powinno to być dokładniejsze 1

(4) Ogólne zdjęcie rentgenowskie ustali kilka rodzajów krzywych, powszechnie używanych jest pięć: PE,PCV, lutować, stop miedzi, aluminum alloy. W ten sposób, jest to wygodne do wykrycia.

54. Jaka jest zasada doboru napięcia i prądu w analizie widma fluorescencji edx? Na jakiej zasadzie?

Ciężkie pierwiastki wybierają duże napięcie i mały prąd, elementy lekkie wybierają ogólnie małe napięcie i duży prąd, metoda testowa linii pomiarowej, której używamy, to domyślne 10 MM dla apertury, ale w rzeczywistości, istnieje wiele próbek mniejszych niż 10 MM, ale inżynier powiedział, że użycie innej apertury do wykonania krzywej standardowej doprowadzi do niedokładnej absorpcji tła, jak mam to wyjaśnić? Tylko w EDX8000B PVC-PE zastosowano aperturę 10 MM, metalem jest zastosowanie apertury 3 MM, więc nie musisz testować średnicy 10 MM podczas testowania metalu. A PVC-PE może ogólnie spełniać wymagania 10 MM. Jeśli naprawdę nie możesz spełnić tego wymogu, to nie ma znaczenia, you first meet the thickness requirements, put the test sample in the middle of the aperture,EDX-8000B will automatically detect the size of the sample and automatically adjust the test area. But it should be noted here that the aperture of the PVC-PE measuring line can not be adjusted to other sizes manually, which will make a large error. The measuring line of metal does not have this function

.If it was going to automatically adjust the aperture size, why would it provide us with a choice of 1,3, 5, I 10?

W rzeczywistości, when you test PVC, if the area of your test sample is not large enough, the aperture will not be adjusted, but the blank place will be removed in the calculation, and the area of this part will not be calculated, so as not to affect the test results. Tam są 4 rodzaje apertur dla Twojej wygody podczas wykonywania testu, ponieważ im większa apertura, tym lepsza dokładność testu. Na przykład, jeśli twoja próbka metalu ma tylko 1 MM, wtedy możesz dokładniej przetestować przy aperturze 1 MM niż przy aperturze 3 MM. Jeśli masz wątpliwości co do wyniku badania próbki metalu, możesz użyć apertury 5 MM lub 10 MM, aby sprawdzić dokładniejszą wartość testową. Metalowa apertura 3MM spełnia jedynie najbardziej podstawowe wymagania, nie jest to najlepsza przysłona testowa.

56. Posiadamy analizator fluorescencji EDX, który jest używany od dwóch lat. Chcesz wiedzieć, czy wynik (PPM) mierzone metodą EDX jest bezpośrednio powiązane z wysokością piku widmowego, czy też jest to bezpośrednio związane z obszarem całego piku widmowego? Jaki jest konkretny wzór konwersji pomiędzy PPM a intensywnością piku widmowego?

(1) Ogólny spektralny wielopikowy obszar chromatograficzny z wieloma pikami dla określonej zależności, którą możesz sprawdzić w pliku pomocy lub instrukcjach instrumentu (powinno być powiązane z licznikiem)

(2) Istnieje bezpośredni związek z tym obszarem

Intensywność – jest ogólnie stosunkiem powierzchni do czasu

Czas ma czas martwy, czas na żywo, a czas rzeczywisty to czas realny

Obliczenie obszaru jest w zasadzie sumą liczby obszarów zainteresowania, jeśli masz podkrzywą y=k(x-b), y to intensywność, a x to PPM

(3) Wyniki mojej analizy pliku danych EDX8000B są następujące:

1. Urządzenie wykorzystuje cps/uA, który uzyskuje się dzieląc cps przez czas detekcji (Na żywo);

2, przede wszystkim sposób montażu, the data points in the detection interval curve fitting, and the curve in this interval integral, to jest, calculate the peak area in the detection area, if you choose the internal standard correction, but also divide the corresponding value of the internal standard object (may be the background), and finally get the Net intensity in the result display (Net Int.);

3, this strength is compared with the strength in the standard curve, and the ppm value of the test sample is obtained.

X-ray fluorescence Spectrometer FAQ Series 2

57. Listen to others say that because sometimes it is not correct to look at the XRF test results (quantitative results) alone, it is necessary to look at the wave crest. Na przykład, Pb, if the peak strength is relatively large, then ICP will be used to further measure the Pb content. Now the question is how to divide the intensity of this crest 1 na duże i małe?

(1) Aby zobaczyć więcej niż dwie linie widmowe, takie jak Pb, musisz spojrzeć na PbLa i PbLb1. Jeśli element Pb naprawdę zawiera, dwie linie widmowe będą miały bardziej wyraźny grzebień, jeśli element Pb rzeczywiście zawiera inne elementy i jest przez nie zakłócany oraz spowodowany przez duże, wówczas dwie linie widmowe będą miały grzbiet, ale wyniki dwóch testowanych linii widmowych będą stosunkowo różne.

(2) Nawet jeśli CdKa(Cd pojedynczego analitu charakterystycznego dla promieni rentgenowskich Ka) ma pewien szczyt, ale inne linie widmowe nie osiągają wartości szczytowych, nie można uznać, że musi zawierać tę substancję. Zazwyczaj, muszą istnieć więcej niż dwie linie widmowe (w tym dwa) w celu określenia piku zawartego w tym elemencie, nie można wyznaczyć linii widmowej.

(3) Jeśli stosowana jest metoda linii dozorowej, metoda oznaczania również powinna być podobna, oznaczanie Pb wygląda następująco, w zależności od tego, czy herb La i LB istnieje w tym samym czasie. Jeśli istnieje tylko jeden herb, nie można stwierdzić, że musi istnieć. Ponadto, aby sprawdzić, czy element zakłócający Pb istnieje, oprogramowanie może wyświetlić przebieg elementu zakłócającego, w przypadku połknięcia grzbietu elementu zakłócającego i grzbietu Pb. Konieczne jest określenie wielkości ciężkich pików, które można rozróżnić i oddzielić programowo.59. Jaką tabletkarkę powinna wybrać firma EDXRF?Tabletkarka jest urządzeniem uniwersalnym, głównym wyborem prostej obsługi, trwała skóra, ciśnienie może spełnić Twoją próbkę, konfigurację zgodną z formą instrumentu, najlepszy wybór elektryczny.

59. Jaką tabletkarkę powinna wybrać firma EDXRF?

Tabletkarka jest urządzeniem uniwersalnym, głównym wyborem prostej obsługi, trwała skóra, ciśnienie może spełnić Twoją próbkę, konfigurację zgodną z formą instrumentu, najlepszy wybór elektryczny.

60. Teraz używam EDXRF, aby uzyskać lepsze wyniki testów RoHS dla tworzyw sztucznych niż dla metali. Pomiary ilościowe metali zawsze są niedokładne, i odchylenie jest duże, szczególnie CD i PB. Czy mogę zapytać, czy krzywa wzorcowa nie jest dobrze wykonana lub jest problem z przygotowaniem próbki, lub ograniczenie samego instrumentu?

(1) W pomiarze metalu występuje więcej elementów zakłócających, co powinno być jedną z przyczyn dużego odchylenia

(2) Matryca jest powodem wpływającym na stabilność analizy. Plastikowa matryca to CHO, co ma niewielki wpływ na interferencję powiązanych elementów analizy. Metal jest duży. Jakość lub dostępność próbek standardowych również wpływa na dokładność analizy.

61. Ostatnio w instrumencie EDXRF do wykrywania produktów galwanicznych, measured the Pb content in the same part of the different measuring points are 300~2800PPM,(metoda: qualitative and semi-quantitative; Próbka :SUM24L fast-cutting steel plating Ni 4-6um; Wipe each part with alcohol on the surface to measure three points), why there is such a big gap, relevant comparative tests have been done, or can not find the problem.B: According to what you said, it should be the X-ray that penetrates the plating. And the coating is not uniform, so there is a thin coating with high lead content, and vice versa

62. The measuring range of X fluorescence spectrometer is generally from which number to which number of elements?

(1) Obecnie, the best energy dispersion spectrometer starts with sodium

(2) wavelength dispersion type, generally from Na to U, a good point (depending on the spectroscopic crystal) can Be from B to U, teoretycznie z być do U, ale generalnie bycie jest trudne.

(3) Kształt długości fali to NA-U, kształt energii powinien móc analizować B E-U, Do poznania używano dwóch rodzajów instrumentów.

64. Jaka jest powtarzalność spektrometru fluorescencji rentgenowskiej?

(1) Mówiąc najprościej, mierzysz tę samą próbkę wiele razy w tych samych warunkach. Czy widzisz, jak odległe są wyniki?? Mały oznacza powtarzalny. W przeciwnym razie, powtarzalność jest słaba. Powtarzalność jest ważną miarą w XRF.

(2) Można to interpretować dosłownie: jest to wskaźnik wydajności powtarzalnych operacji

(3) W metrologii nie ma słowa o powtarzalności, tylko słowo powtarzalność i powtarzalność. Powtarzalność odnosi się do spójności wyników uzyskanych poprzez wielokrotne pomiary tego samego mierzonego obiektu w tych samych warunkach pomiarowych, w tym: ta sama procedura pomiarowa, ten sam obserwator, tego samego przyrządu pomiarowego w tych samych warunkach, ta sama lokalizacja, powtarzane pomiary w krótkim czasie. Powtarzalność odnosi się do spójności tego samego wyniku pomiaru w zmienionych warunkach pomiaru.

X-ray fluorescence Spectrometer FAQ Series 2

69. Czy spektrometr fluorescencyjny X może wytwarzać solankę?

(1) Jeśli na wyposażeniu znajduje się urządzenie do analizy próbek cieczy, można to zmierzyć bezpośrednio; Jeśli nie, można zastosować metody pośrednie, takie jak adsorpcja próbki na bibule filtracyjnej w celu pomiaru i tak dalej

(2) Najpierw jest standard, solanka ogólna K, Już, Ok, Mg jest stosunkowo wysoki, rozcieńczyć. Włóż do plastikowego kubka, w dolnej części znajduje się kilka mikronów folii z tworzywa sztucznego, i użyj promieniowania, nie ma potrzeby odkurzania XRF

(3) Istnieją dwie metody:1 Pomiar odbywa się za pomocą kubka na próbkę cieczy.2 Pomiar odbywa się za pomocą bibuły filtracyjnej.