Opis

GA-MN60D Dual Purpose Gloss Meter

Wstęp
The technical parameters of MN60-D mirror gloss meter conform to the national standard GB9754-88, GB9966.5 and international standard ISO2813. each
All the performance indexes meet the requirements of national JJG696-2002 (Calibration regulation of mirror gloss meter).

This product is suitable for the measurement of gloss of paint, farba, atrament, Plastikowy, stone, papier, tile, enamel and other flat products.

 

GA-MN60D Dual Purpose Gloss Meter

Funkcja

1. The small design can be put in the pocket, more convenient to carry, is the current domestic small volume, light weight gloss meter.
2. Ultra-power-saving design, using a No. 5 bateria (rechargeable battery or alkaline battery can be used), you can operate. A new battery can last for 60 godziny
On it, about ten thousand data points can be measured.
3. Special moisture-proof process allows you to rest assured that it can be used in a variety of harsh environments.
4. Immediately open and test, no waiting time, stable value, no drift, easy to quickly measure.
5. Long life light source, no need to replace.
6. The matching between the light source and the photoelectric conversion meets the SD65 (L) ω condition.
Seven. Bluetooth BLE communication, can be connected with the mobile phone test, and the use of the machine is no different.
8. Adapt to MyGloss APP for Android phone to realize multi-functional data processing such as storage statistics.

 

GA-MN60D Dual Purpose Gloss Meter

Notatka:
1. Avoid direct light during use, otherwise the measurement accuracy will be affected.
2. The standard board of the instrument should be kept clean. Do not touch the surface with your fingers. If the surface is stained, it will lead to increased instrument error and inaccurate reading
Wipe with head paper or anhydrous ethanol.
3. Because of the high stability of the instrument, the instrument does not need to be calibrated frequently, it is recommended to re-calibrate once after continuous use for 2 godziny.
4, the battery should be replaced in time, otherwise long-term storage, the old battery may flow out of waste liquid, causing damage to the instrument.

 

GA-MN60D Dual Purpose Gloss Meter

Parametr

skala 0~99.9Gs/100-1000Gs (Gloss unit)
stability Less than ± 0.4Gs /30Min
indication error 0-99.9<±1.5Gs/100-1000<±2Gs
zasilacz roboczy One 5 (AA) alkaline battery or rechargeable battery
working voltage 0.8V~1.5V
Beam Angle 60°
rozmiar miejsca 11x22mm
Measuring port size 14x28mm
service conditions environment temperature :0℃~40℃
Wilgotność względna: nie więcej niż 85%
Main engine size (L)114mm x (W)32mm x (H)64mm
Standard board box size (długość)67mm x (szerokość)46mm x (thickness)13mm
Main engine weight 300G
Standard box weight 45G
Software adaptation My Gloss APP: 2.0(Android 4.1 powyżej)