Beschrijving

GA-MNB2 Blackness Meter

Invoering:
GA-MNB2 is a portable reflective blackness meter developed by our company. The reflective blackness meter is a highly sensitive reflectance measuring instrument for measuring black and dark gray objects
The blackness of the body surface.
It is suitable for the determination of the blackness of carbon black in ink coating, carbon black and rubber industry, and the blackness of negatives, plastics, ceramics, papier, enz.

 

GA-MNB2 Blackness Meter

instrument features:
1. The ultra-small design can be put in the pocket, more convenient to carry and field measurement, is the current domestic small volume, light weight blackness meter.
2. Ultra-power-saving design, using a No. 5 battery (rechargeable battery or alkaline battery can be used), you can operate. A new battery can last for 50 uur
On it, about ten thousand data points can be measured.
3. No zero calibration required, eenvoudige bediening.
4. Special moisture-proof process allows you to rest assured in a variety of harsh environments.
5. Long life light source, no need to replace.

 

GA-MNB2 Blackness Meter

Precautions:
1. Avoid direct light during use, otherwise the measurement accuracy will be affected.
2. The standard board of the instrument should be kept clean. Do not touch the surface with your fingers. If the surface is stained, it will lead to increased instrument error and inaccurate reading
Wipe with head paper or anhydrous ethanol.
3. Because of the high stability of the instrument, the instrument does not need to be calibrated frequently, it is recommended to re-calibrate once after continuous use for 2 uur.
4, the battery should be replaced in time, otherwise long-term storage, the old battery may flow out of waste liquid, causing damage to the instrument.

 

GA-MNB2 Blackness Meter

Parameter

meetbereik 0~199.9Gs
lichtbron D65 light source lighting in accordance with CIE regulations
Measuring Condition 45/0(45 degrees transmit, 0 degrees receive)
zero deviation With automatic zero stabilization function, no need to adjust zero
indication error less than ±1.5Gs
herhaalbaarheid 0.2Gs
Indication drift less than ±0.4Gs/30Min
Meetnauwkeurigheid 0.1
werkende voeding One 5 (AA) alkaline battery or rechargeable battery
werkspanning 0.8V~1.5V
Measuring aperture 12×16mm
service conditions environment temperature:0℃~40℃
Relatieve vochtigheid: not more than 85%
Dimensie (L)114mm x (W)32mm x (H)64mm
Main engine weight 300G