Main technical parameters
Measurement density range: 0 ~ 5.0D;
Measure field: ф2mm(diameter of light hole);
Herhaalbaarheidsfout: ±0.03d;
Meetfout :(0.00d ~ 3.00d)±0.03D (3.00 ~ 4.00 D)±0.04D; (4.00 ~ 5.00 D) ±0.06D.
Weergavemodus: three-digit LED digital display;
Stroomvoorziening: AC220V + + 10% 50HZ;
Power consumption: less than or equal to 7W;
8using the environment: 0 ~ 40 ℃, relatieve vochtigheid 85% or less;
External dimensions: 330 X 110 x 116mm;
Gewicht: 1.7kg.
GA-3813 Zwart-wit densimeter
Working principle
GA-3813 digital transmission black and white densitometer is used to measure the amount of light passing through transparent materials (such as film base) to determine the amount of light absorbed by the surface of the instrument, it is mainly composed of lighting system, acquisition of light and measurement system and signal processing system.Ac modulation light source is used to provide a light source with certain quality and stability. The amount of light passing through the film under test is received by the photoelectric sensor and converted into electrical signals. After signal processing, the light is sent to the microcontroller and the density value of the film under test is finally displayed on the display.
GA-3813 Zwart-wit densimeter
Configuration sheet
black and white densitometer host: one
Stroomdraad: one
Standard density film: one piece
Spare fuse: one
Instruction manual: one copy
Instrument warranty card: one copy
Certificate of qualification: one copy