GA-1019 rubberen Mooney-viscositeitsmeter
- Beschrijving
- Navraag
Beschrijving
GA-1019 rubberen Mooney-viscositeitsmeter
Test standard:
Accord with GB/T1232″Rubber unvulcanized -Determination of mooney viscosity”,GB/T 1233 ”Rubber,unvulcanized-Determination of prevulcanization characteristics-Shearing disk method”,ISO289,ISO667ect.
GA-1019 rubberen Mooney-viscositeitsmeter
Samenvatting:
Mooney Viscometer is used for measuring viscosity and relaxation of base rubber, raw rubber stock or rubber mass prevulcanization properties, it is also used for measuring viscosity in Mooney value for compounded or uncompounded unvulcanized natural, synthetic or regenerated rubber.
This instrument has excellent performances including automatic calibration, higher stability. It can perform the following tests: viscosity, scorch and stress relaxation.
GA-1019 rubberen Mooney-viscositeitsmeter
Functies:
- Easy to use the instrument.
GA-1019 have automatic door and all of the electric components are CE corresponding.
- Temperature can be increased up to 185 within 7.5min, and recovery time when beginning, and recovery time when beginning test will be less 2min. And temperature fluctuation will be within 0.1.
- All instruments adopted the way of PID to control temperature. The smallest size and weight in all of the same type instrument in the market;
- The die was designed to international standard ISO.
- With lower noise when working, ≤55dB.
- It is following ISO 289 ,ASTM D 1646, JIS K 6300 .
- Can variable Temperature unit “C, “ F Degree.
8.Use window XP program connect to machine & and can setting Mooney or scorch time and show Dynamically Mooney value or scorch time on the graph.
GA-1019 rubberen Mooney-viscositeitsmeter
Specificatie:
| Range of temperature control
|
From room temperature to 200°C |
| Temperature fluctuation in die | ≤±0.1°C |
| Temperature display resolution | ± 0,1 ℃(50-150℃) |
| Temperature stabillity
|
± 0,1 ℃(50-150℃)
(50-150℃) |
| Rate of temperature increase | 25°C/min |
| Setting range | 0-150℃;Min unit:0.1℃
|
| Setting time range | 0-200minuten |
| Temperature display range
|
0-150℃,display resolution: 0.01℃ |
| Display range switch | Automatic increase and decrease according to actual torque (0-200 menni) |
| Mooney value display accuracy | 0.01Mooney value |
| Stroom | 220V±10%,50±1Hz,0.8kW,It must be well grounded |
| Compressed air | 0.4Mpa,Controlled by pneumatic control valve |
| Rotorsnelheid | Fixed 2 times/min |
| Cavity type | Rectangular deep groove die cavity |
| Environment temperature | Normal temperature -40℃, relatieve vochtigheid < 80% (no corrosive gas in the air) |
| Gewicht | 160kg
|
| Outside size | 650*650*1100mm
|
| Measurement range of torque | 0-200 MV |
| Torque resolution | 0.01MV |
| Frequency of rotor | 2±0.02r/min |
| Temperature PID Control | +/- 0.1°C |
| Upper & Lower | Separate plate temperature control |
| Stroomvoorziening | 220 V ,1 Phase, 50 Hz |