Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
An introduction
Film thickness meter EDX8000T Plus product features
The new downshot integrated design
Fast testing, tiada penyediaan sampel diperlukan
Kamera HD CCD terbina dalam boleh digunakan untuk memerhati dan memilih ukuran ketebalan salutan di kawasan kecil untuk mengelakkan sentuhan langsung, pencemaran atau kerosakan pada objek yang diukur. The software is equipped with distance correction algorithm to realize the difference of irregular samples (seperti permukaan cekung dan cembung, benang, permukaan melengkung, dan lain-lain.)Accurate testing of the profile
Pelbagai sampel standard tersedia untuk pengukuran ketebalan salutan dan analisis komposisi
Can cover the periodic table Mg magnesium to U uranium
Pengesan SDD dengan julat pengiraan tinggi dan resolusi tenaga yang sangat baik
Automatic switching collimator and filter (0.15mm,0.2mm,0.3mm)
Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
XRF coating thickness gauge working principle
The EDX8000T Plus is a coating thickness gauge that irradiates X-rays onto the sample and measures the thickness of metal films such as the coating by the strength of the second X-ray reflected off the sample, which will not cause damage to the sample because it is not in contact with the sample and the X-ray energy that is irradiated on the sample is very low. Pada masa yang sama, pengukuran juga boleh diselesaikan dalam 10-30 detik.
Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
Typical application directions:
Coating thickness analysis
Penyaduran krom, seperti produk plastik dengan kemasan penyaduran krom hiasan
Salutan antikarat seperti zink pada keluli
Salutan pada papan litar dan PCBS fleksibel
Contact surfaces for plugs and electrical contacts
Plating solution analysis
Analysis of rhodium materials on precious metal coatings such as gold substrates
Plating solution analysis
Analisis salutan berfungsi untuk industri elektronik dan semikonduktor
Analyze hard material coatings such as CrN,TIN or TiCN2 Instrumentation and application technical requirements
Alat: EDX8000T PLUS X-ray fluorescence spectrometer
Application information: The substrate can be plastic or metal. The coating material must be metal. The coating can be pure metal or alloy. Pure metal can be used to establish new calibrations for different coating materials. Software based on the basic parameter algorithm (FP program) allows the instrument to measure not only the coating thickness of each layer, but also the elemental chemical content of each layer.
Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
Measuring parameters
| Tube voltage | Tube current | Penapis | Collimator | Masa ujian |
| 40kv | 600uA | AI | 0.5mm | 100s |
Three testable maximum number of layers and error analysis:
1. maksimum 5 layers at the top of the substrate. Chemical composition analysis for each layer ranges from 5ppm to 100%
2. Detection limit of coating thickness: 0.01μm
3. Ketepatan: from top layer to bottom layer
The first layer is ±5%
The second layer is ±10%
The third layer is ±20%
The fourth layer is ±25%
The fifth layer is ±30%
Penetration depth of EDXRF using a 50 watt X-ray tube
Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
| XRF penetrable thickness gauge | ||
| Coating material | Base material | Ketebalan (μml) |
| AI | Cu | 0-100.0 |
| cd | Fe | 0-60.0 |
| cu | Al | 0-30.0 |
| cu | Fe | 0-30.0 |
| cu | Plastic | 0-30.0 |
| Au | Ceramic | 0-8.0 |
| Au | Cu/Ni | 0-8.0 |
| pb | Cu/Ni | 1-15.0 |
| Dalam | Ceramic | 0-20.0 |
| Dalam | Cu | 0-20.0 |
| Dalam | Fe | 0-20.0 |
| pd | Dalam | 0-40.0 |
| pd-Ni | Dalam | 0-20.0 |
| pt | Daripada | 0-8.0 |
| Rh | Cu/Ni | 0-50.0 |
| Ag | Cu/Ni | 0-50.0 |
| Sn | Al | 0-60.0 |
| Sn | Cu/Ni | 0-60.0 |
| Sn-pb | Cu/Ni | 0-25.0 |
| Zn | Fe | 0-40.0 |
Four copper nickel tin plating spectrum and test results are as follows
The spectrum and test results of iron, copper and tin plating are as follows
Coating thickness analysis solution –EDX800OT PLUST PLUS spectrometer application report
5. Customizable test reportsSample type Copper gold plating Three consecutive test reports: base copper, plating pure gold
Six conclusionsAccording to the above coating test performance and report, it can be concluded that EDX-8000T PLUS series XRF can be used to analyze the coating thickness of different materials and identify the chemical composition content of materials at the same time.