Keterangan

Oxygen Analyzer

1.Public environmental condition

Nomor Barang Environmental conditions Catatan
1 service voltage 220V ± 10%,50±2%Hz  
2 resistensi tanah <5Oh(Separate ground cable required)  
3 kelembaban <80%RH,An air conditioner is required to be installed in the operation room  
4 environment temperature 18℃~29℃  
5 electromagnetic interference, vibration No high frequency electromagnetic field, no strong vibration source  
6 carrier gas Argon, purity 99.999%  
7 power gas

 

Nitrogen, industrial pure  

2.Desain, manufacture, installation and acceptance of legal standards and standards

Nomor Legal specification standard name

 

 

Catatan
1 GB/T 11261-2006 Determination of oxygen content of iron and steel – pulse heating inert gas melting – infrared absorption method

 

 

3.Equipment and supporting materials purchase list

Nomor Nama Factory owner

 

 

Quantity

 

 

Catatan
1 Oxygen analyzer main engine

 

 

  1 bagian konfigurasi standar
2 Standard spare parts are provided

 

 

For details, see the spare parts list 1 mengatur konfigurasi standar
3 analysis software   1 mengatur konfigurasi standar
4 Instrument instruction manual   a copy konfigurasi standar
5 Cooling circulation system 2 HP cooling capacity 1 mengatur konfigurasi standar
6 pencetak HP printer 1 bagian konfigurasi standar
7 electronic scales Sartorius 1/10,000 ketepatan 1 bagian konfigurasi standar
8 computer Industrial Personal Computer 1 mengatur konfigurasi standar
9 voltage-stabilized source 10KVA 1 bagian konfigurasi standar
10 standard sample One set according to customer needs 1 mengatur konfigurasi standar

4.Technical requirements

1、Major parts requirements

Nomor Nama Brand or technical requirements Catatan
1 detecting system German brand konfigurasi standar
2 pneumatic element Italian brand konfigurasi standar
3  

Pulse furnace

 

water-cooled transformer konfigurasi standar
4 linear power supply Aerospace Science and Industry Group Chaoyang Changfeng konfigurasi standar
5 temperature control system Taiwan brand konfigurasi standar
6 power controller Taiwan brand konfigurasi standar

2、Equipment technical specifications requirements

Nomor Barang require Catatan
1 Principle and application The oxygen in metal and non-metal solid materials is detected by inert gas melting principle. In the determination of oxygen, the sample after weighing is placed in a graphite crucible and melted by heating at high temperature in the argon gas flow. The oxygen in the sample reacts with the carbon in the graphite crucible to produce carbon monoxide (CO). This mixture is sent by argon gas to the reformer, where the carbon monoxide (CO) is converted to carbon dioxide (CO2), and through the gas mixture after the reformer is sent to the infrared detection tank, where carbon dioxide (CO2) is detected. Rapid and accurate determination of oxygen content in steel, cast iron, alloys, tembaga, zirconium, titanium, rare earth materials, nanomaterials and other inorganic substances.

 

 

 
2 analytical range Oxygen: 0.00001% ~ 20%(expandable measurement range)

 

 

 
3 analytical precision Oxygen: 1ppm or RSD≤1%

Catatan: not greater than the sample standard deviation or uncertainty shall prevail.

 

 

 
4 kepekaan 0.01ppm  
5 parsing time Biasanya 3 menit

 

 

 
6 detecting system Non-dispersive infrared detection system was used for oxygen analysis

 

 
7 Infrared detection system 1, the standard configuration oxygen analyzer is equipped with an independent infrared absorption pool, and the length and number of channels can be flexibly configured according to user needs;

2, motor: the modulation system adopts single-chip microcomputer control of high-precision stepper motor, to achieve long-term stability of the modulation frequency;
3, sumber cahaya: infrared light source selection of efficient, long-life precious metal micro-infrared luminescent and gold-plated metal mirror;
4, constant temperature: the whole gas chamber for constant temperature control, to ensure the analysis of constant temperature, to ensure the measurement accuracy, the use of temperature compensation semiconductor detector, to ensure that the instrument by the outside influence is small, the baseline is very stable.

 

 

 
8 Sample weighing 0.01-1G,Sample size can be changed according to sample content

 

 
9 Pulse furnace

 

1, saat ini 0 ~ 1500A, kekuatan: 10KVA, the highest temperature is higher than 3200℃.

2, pulse electrode furnace heating temperature is high, using program control power, can provide constant power heating, slope heating, segmented heating and other program heating methods, improve the analysis accuracy. It is suitable for rapid analysis from low melting point materials such as aluminum alloy to high melting point materials such as tungsten alloy. The electrode is suitable for a variety of types of crucible, according to the release of different samples, can choose a variety of crucible including sleeve crucible, high temperature crucible, effectively reduce the user’s use cost. The electrode cooling system adopts a single cycle independent cooling system, which can directly connect the circulating water or the external water cooling device, and has the characteristics of good cooling effect and high reliability.

 

 

 
10 gas circuit 1, the use of integrated whole machine design, open space, the gas path system connection route is greatly reduced, the whole gas path system is more concise and reliable. Gas components including solenoid valve, cylinder, gas pipe, gas joint all adopt imported components, solenoid valve life of more than one million times. The special lifting system with two-axis guide cylinder ensures the smooth lifting of the electrode, ensures the good contact between the graphite crucible and the upper electrode, and ensures the uniform force of the graphite crucible, which is conducive to long-term high temperature heating. Automatic blanking furnace head with air curtain protection and automatic air purging mechanism effectively ensure the accuracy of oxygen element analysis. The furnace head is a split structure, easy to disassemble and replace, and the lower electrode is made of high-temperature alloy material, long service life and low cost. (If the whole furnace head is used, the replacement of the electrode requires professional and technical personnel to operate, which is very troublesome).

2, the design of the sampling system is open, its advantage is that the pollutants are discharged from the top of the furnace, not into the system, mudah dioperasikan, the system sealing requirements are not high.(Closed, the gas path through a costly six-way valve to control, internal out.(Some companies have complex sampling device, there are two cylinders, such as the sample stuck, need to open the cylinder, difficult to take out). The micro-flow sensor is used to achieve high precision flow control, and the influence of analysis air flow changes on the analysis is minimized. Gas conversion through this valve, so that the valve is easy to break, dust and other pollutants are also easy to bring into the system). This injection device is simple and can be automatically injected, in case the sample is stuck, easy to remove

 

 
11 Power gas

 

Nitrogen or water-free oil-free compressed air

 

 

 
12 revise 1. Single point correction: correction with a single analysis result

2, multi-point correction: with multiple analysis results correction
3. Blank correction

 

 

 
13 catu daya 1, pulse furnace power supply: 220VAC 10%, 501Hz, maximum current 50A.

2. Other power supply of the instrument: 220VAC 10%, 501Hz, maximum current 10A.
3, circulating water power supply: 220VAC 10%, 501Hz, maximum current 16A.

 

 

 
14 fungsi perangkat lunak 1. Provide software under Windows environment to quickly display analysis results and analysis curves;

2. Automatic storage of analysis results;
3. The sample weight is automatically entered into the software;
4. Channel setting is optional;
5, dynamic release curve drawing and storage, unloading;
6, the analysis results of a variety of correction modes, can be manually corrected, can also be corrected through the analysis results; Can be single point correction, can also be multi-point correction;
7, the data screening, average, standard deviation, relative deviation, dll.;
8, the new software can be based on user application, can design a number of calibration lines, greatly facilitate the use of users;
9, the host can and computer, pencetak, electronic day equal connection to achieve data transmission;
10. The analysis time and blank value can be adjusted, and the blank value of the melting agent can be automatically removed;
11, the user can store several analysis methods for different analysis samples.
12, can store a lot of sample weight (quantity is limited only by the computer hard disk capacity).
13, can output regular format results report.
14, can be connected with the user database, to achieve remote transmission of analysis results, easy to achieve the network management of analysis results.
15, can quickly carry out instrument blank determination.
16, sample number can be stored and edited.
17, can be applied to automatic sampling, to achieve automatic operation of the instrument.
18, oxygen can be arbitrarily set up high and low analysis channel to meet the analysis requirements of different content and different materials. Users can use the combination freely according to actual needs.

 

 

 
15 routine test  

Instrument inspection and calibration certificate

 

 

 

5.Provide equipment data requirements

Nomor Nama Catatan
1 manufacturer certificate  
2 Operating standards for equipment operation, inspection and maintenance  
3 Spare parts list of the whole set  
4 Machine operation manual.  
5 Operating instructions for main parts.  

6.Daftar suku cadang

6A-101 Graphite external crucible 100 Bagian-bagian    
6A-102 Graphite internal crucible 300 Bagian-bagian    
6A-103 decompressing table 2 Bagian-bagian    
6A-104 hairbrush 1 Bagian    
6A-105 instruction book 1 Bagian    
6A-106 cleanroom wiper 1 Bagian    
6A-107 High temperature electrode sheet 1 Bagian    
6A-108 Seal spare parts 1 Mengatur    
6A-109 Sample tweezers 1 Bagian    
6A-110 Crucible tweezers 1 Bagian    
6A-111 Pagoda copper brush 1 Bagian    
6A-112 Injector cleaning brush 1 Bagian    
6A-113 air pipe line Φ6 20 Meters    
6A-114 Vacuum silicone grease 1 Bagian    
6A-115 High efficiency dehydrating agent 1 Bagian    
6A-116 CO2absorbent 1 Bagian    
6A-117 quartz wool 1 Bagian    
6A-118 water pipe 20 Meters    
6A-119 Rare earth copper oxide 1 Bagian    
6A-120 High temperature quartz tube 1 Bagian    
6A-121 Reagent quartz tube 1 Bagian    
6A-122 Filtered quartz tube 1 Bagian    
6A-123 certificate 1 Bagian    
6A-124 Nickel basket 1 Bagian    
6A-125 Nickel pocket 1 Bagian    
6A-126 Electrode cleaning copper brush 1 Bagian    

 

Oxygen Analyzer