Description

Integrated Magnetic Analyzer

Caractéristiques
High-resolution: 7-pouce 800 × 480 dots, TFT-LCD display
Six-in-one comprehensive analysis, one machine can achieve the
comprehensive test needs of coil components
High-power AC withstand voltage analysis
DC voltage analysis
Insulation resistance analysis
Turn-to-turn insulation analysis
DC low resistance analysis
Inductance test analysis
Eight-channel switching technology that can test eight different
components simultaneously
500VA power AC withstand voltage design, in line with UL 1004-1 motor
test standards
Test de résistance à l'isolation: maximum voltage can reach 5kV
CC / IR automatic rapid discharge function
Turn-to-turn insulation test: sampling ADC promoting to 12bit, 200MHz
sampling rate
DC low resistance test: support DC resistance calculation of Y-type motor
DC low resistance temperature conversion function and optional
capteur de température
Inductance test analysis of up to 100kHz frequency
Quick contact check function to realize rapid detection of test fixture
New-type high voltage test fixture
Four-terminal Kelvin test of DC low resistance and inductance
Test steps up to 32
Internal file storage and external U disk file saving

 

Integrated Magnetic Analyzer

Paramètre technique

Modèle TH9520
Number of Channels 8
Withstand Test
Output Voltage CA 0.050 – 5.000kV,Step 0.001kV,Frequency 50Hz/60Hz ±0.1%,sinusoidal waveform
CC 0.050 – 6.000kV,Step 0.001kV
Précision ± (1% valeur définie + 0.1% of full scale)
Adjustment Rate (1% output + 0.1% of full scale) rated power
Current Range CA Voltage≤4.000kV: 0.001mA – 120.0mA, Tension>4.000kV:0.001mA – 100.0mA
CC 0.1faire – 20.00mA
Précision ± (1% of reading + 0.5% of full scale), AC Real: ± (1% of reading + 5% of total current reading + 5 chiffres)
Output Power 500Virginie
ARC CA 1.0mA – 20.0mA,0.1mA Step
CC 1.0mA – 10.0mA,0.1mA Step
Insulation Resistance Test
Output Voltage 0.050 – 5.000kV,Step 0.001kV Accuracy: ± (1% of set value + 0.1% of full scale)
Resistance Test Range 0.100MΩ- 99.99GΩ Resolution:0.1MΩ
Measurement Accuracy ≥500V 1.000MΩ – 1.000GΩ, ± (3% of reading + 5 chiffres)
1.000GΩ – 10.00GΩ, ± (7% of reading + 5 chiffres)
10.00GΩ – 99.99GΩ, ± (10% of reading + 5 chiffres)
< 500V 0.100MΩ – 1.000GΩ, ± (7% of reading + 5 chiffres)
1.000GΩ – 99.99GΩ,pour référence seulement, no accuracy requirements
Time Setting
Rise Time OFF,0.1s – 999.9s,Step 0.1s
Test Time 0.1s – 999.9s,Step 0.1s
Fall Time OFF,0.1s – 999.9s,Step 0.1s
Waiting Time OFF,0.1s – 999.9s,Step 0.1s
Turn-to-turn Insulation Test
Output Pulse Voltage 0.01kV – 6.000kV,0.01kV Step,± 5% set value ± 15V
Inductance Test Range ≥10μH
Pulse Energy jusqu'à 0.36 Joule
Waveform Sampling Taux d'échantillonnage: 12bit, Vitesse d'échantillonnage: 200MHz, adjustable 8-level, Memory depth: 12k Byte, Sample average: 1 – 32
Number of Applied Pulses jusqu'à 32
Judgment Method Area comparison, area difference comparison, corona discharge, phase difference comparison
Low Resistance Test / and Y Type Resistance Test
Test Signal mΩ 1A,1Ω 0.5A,others ≤3V
Plage de test 0.01mΩ – 1.2MΩ
Résistance Gamme 0.01mΩ – 120.00mΩ 0.1mΩ – 1200.0mΩ 0.001Oh – 12.000Oh 0.01Oh – 120.00kΩ 0.1kΩ – 1200.0kΩ
Précision ± 0.5% of reading +0.04% of full scale ± 0.3% of reading +0.03% of full scale ± 0.2% of reading +0.03% of full scale ± 0.1% of reading +0.03% of full scale ± 0.2% + 0.03% of full scale)
Inductance Test
Test Parameters Ls,Lp,Rs,Rp,Q
Measurement Accuracy 0.5%
Test Frequency 100Hz,120Hz,1kHz, 10kHz, 100kHz
Test Signal Level 1.0Vrms,10% précision

Integrated Magnetic Analyzer Integrated Magnetic Analyzer Integrated Magnetic Analyzer