Description

Hipot Tester

Caractéristiques
TH9310 series: CA:5kV/10mA; CC:6kV/5mA AC/ DC withstanding
voltage/insulation resistance tester
TH9320 series: CA:5kV/20mA; DC6kV/10mA AC/ DC withstanding
voltage/insulation resistance tester
TH9310/20:AC/ DC withstanding voltage/insulation resistance tester
TH9310A/20A: CA / DC withstanding voltage tester
TH9310B/20B: AC withstanding voltage tester
480×272 dot-matrix, TFT-LCD display
Rapidly discharging and arc detection
Randomly set voltage rising time and testing time in 999.9 secondes;
Freely set waiting time for insulation resistance
Hold 5 testing steps; 4 testing modes selectable
Brand new operation interface and concise interface operation design
Lock keyboard
PLC interface

Hipot Tester

Paramètre technique

Modèle TH9310/20 TH9310B
Withstanding Voltage Test
Output Voltage CA 0.05 —5.00kV ± (2% reading+5digits), (50Hz,60Hz optional)
CC 0.05 —6.00kV ± (2% reading+5digits) ——
Voltage Adjustment Rate ≤ (1% – 5V)(rated power)
Current TestRange CA TH9310:0.000mA – 10.00mA ± (2% reading+2digits)TH9320:0.000mA – 20.00mA ± (2% reading+2digits)
CC TH9310:0uA – 5.00mA ± (2% reading+2digits)TH9320:0uA –10.00mA ± (2% reading+2digits) ——
Discharge Function Discharge after test ends(DCW)
Insulation Resistance Test
Output Voltage 0.10kV – 1.00kV ± (2%reading+2V) ——
Resistance Test Range 1MΩ– 9999MΩ ——
Resistance Test Accuracy 500V-1000V 1MΩ– 1000MΩ ± (5%reading+2digits)1000MΩ–9999MΩ ± (10%reading+2digits) ——
100V-500V 1MΩ– 1000MΩ ± (10%reading+2digits) ——
Discharge Function Discharge after test ends ——
Arc Detection
Plage de mesure CA 1 – 9 levels (factory default 5)(20mA,18mA, 16mA, 14mA, 12mA, 10mA, 7.7mA, 5.5mA, 2.8mA respectively)
CC 1 – 9 levels ——
General Specification
Memory 5 groupes
Voltage Rising Time 0.1s – 999.9s
Test Time Setting(AC/DC) 0.2s – 999.9s
Waiting Time(ET) 0.2s – 999.9s ——
Time Accuracy ±(1%+0.1s)
Dimension(W×H×D) 280mm×89mm×428mm/10kg
Interface
Standard HANDLER, RS232, USBDRV(PC interface), USBHOST(USB port)