Testeur d'indice d'oxygène GA-9001-B
- Description
- Enquête
Description
Testeur d'indice d'oxygène GA-9001-B
Design standard:
GB/T 2406.2-2009 GB/T 2406.1-2008.ASTM D 2863, OIN 4589-2, NES 714 GB/T 5454 GB/T 10707 GB/T 8924-2005 GB/T 16581-1996 NB/SH/T 0815-2010 TB/T 2919-1998 CEI 61144-1992 OIN 15705-2002 OIN 4589-2-1996
Testeur d'indice d'oxygène GA-9001-B
Scope of application:
It is suitable for the determination of the combustion performance of plastics, caoutchouc, fiber, foam plastics, film and textile materials
Testeur d'indice d'oxygène GA-9001-B
Principaux paramètres techniques
- Using imported oxygen sensor, Concentration d'oxygène à affichage numérique sans calcul, more accurate and more accurate, gamme 0-100%
- The built-in temperature controller to control the temperature, the control range: 23 ℃ ±2 ℃
- Digital resolution: 0.1%
- Précision de mesure: 0.4 grade
- Plage de régulation de débit: 0-10l/min (60-600l/heure)
- Response time: < 5S
- High temperature resistant quartz glass tube, inner diameter 75 ㎜ height 480 mm or more
- Gas flow rate in combustion cylinder: 40mm ±2mm/s
- Précision du manomètre 2.5, résolution: 0.01mpa
- The flowmeter: 1-15 l/min (60-900 l/H) est réglable, test environment: la température ambiante: température ambiante ~ 40 ℃;, humidité relative: 70%;
- Pression d'entrée: 0.2-0.3mpa
- Pression de travail: nitrogen 0.1-0.15mpa oxygen 0.1-0.15mpa oxygen/nitrogen mixed gas inlet: including pressure regulator valve, flow control valve, gas filter and mixing chamber.
- The sample holder can be used for soft and hard plastics, textiles, fire doors, etc.
- Propane (butane) système de mise à feu, flame length 5mm-60mm can be adjusted freely
- Gaz: nitrogen and oxygen for industrial use with purity of > 99%;(user supplied).
- Exigences de puissance: AC220 (±10%) V, 50Hz
- Maximum working power: 50W
- Igniter: made of a metal, end diameter Φ 2 ± 1 mm nozzle, can ignite sample in combustion tube, la longueur de la flamme: 16 ±4 mm, the size is adjustable
- Porte-échantillon de matériau autoportant: it can be fixed in the position of the axis of combustion cylinder and can hold the sample vertically
- Non-self-supporting material sample holder: two vertical edges of the sample can be fixed on the frame at the same time (optional for non-self-supporting materials such as textiles and films).
Testeur d'indice d'oxygène GA-9001-B
Chassis and part of the structure:
- Control box: CNC machine tool processing molding, electrostatic spraying, beautiful, anti-rust and anti-corrosion.The control part is separately controlled from the test part
- The combustion tube: high temperature resistant high quality quartz glass tube (inner ¢75 mm, 480 mm) long export inner diameter: phi 40 mm
- Sample fixture: self-supporting fixture, and can hold the sample vertically;(optional non-self-supporting style frame), two sets of style clips meet different test requirements;Style clip plug type, style and style clip more simple
- The end of a long pole igniter tube hole diameter ¢2 ± 1 mm, igniter flame length (5 -) 50 mm can be arbitrary adjustable