With powder tabletting method for determination of platinum rhodium and palladium in catalyst – XRF spectrometer Compass200 in ternary catalytic applications

With powder tabletting method for determination of platinum rhodium and palladium in catalyst – XRF spectrometer Compass200 in ternary catalytic applications

(1) sample for the powder particles, pagpapatayo ng mga, grinding, the samples of tablet, then scan test

(2) by XRF method, the algorithm USES the experience coefficient method and the basic parameters of FP/sports schools are

(3) according to the existing fly thoughts of automobile exhaust catalyst secondary sample calibration curve, and then test the sample customer provided

三元催化剂光谱图.jpg

With powder tabletting method for determination of platinum rhodium and palladium in catalyst – XRF spectrometer Compass200 in ternary catalytic applications

20210525

The repeatability test report of the sample (unit: mg/kg)

Sample name Rh Pd Pt
Catalyst 4#-1 35.088 175.501 242.849
Catalyst 4#-2 36.788 167.881 248.069
Catalyst 4#-3 36.063 171.135 244.861
Catalyst 4#-4 36.750 166.669 248.872
Catalyst 4#-5 33.926 177.190 246.872
Catalyst 4#-6 34.488 167.501 250.849
Catalyst 4#-7 35.188 174.281 246.067
Mean value 35.484 171.451 246.920
Standard deviation 1.0982 4.2556 2.6491
Relative standard deviation 3.09% 2.48% 1.07%