GA-2009-A Two Box Thermal Shock Chamber
- Descripción
- Consulta
Descripción
Two Box Thermal Shock Chamber
Especificación
| Nombre del producto | Two van hot and cold impact testing machine
Use of low temperature and high temperature hot and cold storage tank, in accordance with the cylinder, the hanging basket in turn to high temperature and low temperature box, so as to achieve rapid temperature shock effect, balance (BTC) + special temperature control system design of supply air circulation system, to control SSR P.I.D. forma, make the system of heat equal to the amount of heat loss, thus the use of long-term stability. |
|
| Product model | TSD – 11 – a – 2 pag (air-cooled type)
Nota: air-cooled type need not install cooling towers, water cooled type to install, depends on customer’s requirement. |
|
| Limit sample | The test equipment prohibited:
(1)Which is inflammable, explosivo, easy to volatile substances sample test or store (2) the corrosive substance sample test or store (3) biological sample test or store (4) strong electromagnetic emission source sample test or store |
|
| Two Box Thermal Shock Chamber
Volumen, size and weight |
||
| Nominal content product | 52l | |
| Shopping basket size | W50*H32*D32cm | |
| Installation dimensions | W132*H187*D155CM | |
| peso | About 400㎏ | |
| Two Box Thermal Shock Chamber
Actuación |
||
| Test environment conditions | Environment temperature is 25 ℃, relative humidity of 85% or less, test under the condition of no sample in the box
GB/T 5170.2 1996 temperature testing equipment. |
|
| High greenhouse | +60℃ →+150 ℃ | |
| Tiempo | +60℃→+150℃ ≤25min | |
| Rango de temperatura | -55℃~-10℃ | |
| Tiempo | +20℃ → -55℃≤60min | |
| Prueba (style) | Test chamber upside for high temperature, lower for the low temperature zone, specimen is placed in the basket, basket turns in the high temperature box and low temperature in experiment | |
| Shock temperature range | (+60~+100)℃/(-40~-10)℃ | |
| Temperature fluctuation degree | ±0,5 ℃ | |
| The temperature deviation | ± 2.0 ℃ | |
| The temperature recovery time | ≤5min | |
| Temperature conversion time | ≤10sec | |
| Restore conditions | High temperature exposure: + 150 ℃ / mín. – > low temperature exposure: 30 a 55 ℃ / 30 mín.
Sample maximum weight: 2.5 KG |
|
| Meet the test standard | GB/T 2423.1 2008 prueba: A test method at low temperature
GB/T 2423.2 2008 test B: high temperature experiment method GB/T2423.22-2002 test N: temperature change test method test Na GJB 150.5 A – 2009 military equipment laboratory test method for part 5: the temperature impact test GJB 360 a – 1996 temperature impact test GJB367.2-2001 military communication equipment general specification temperature impact test CEI 68-2-14 test method for N: temperature changes |
|
| Two Box Thermal Shock Chamber
Características estructurales |
||
| Thermal insulation enclosure structure | (1) the outer wall material: advanced electrolytic plate surface treatment of the lacquer that bake.
(2) the lining material: stainless steel plate SUS ﹟ 304. (3) low greenhouse body heat preservation material, high temperature resistant rigid polyurethane foam. (4) high greenhouse body heat preservation material: glass fiber high temperature resistant white cotton. (5). The door is thermal insulation material: glass fiber |
|
| Canal de aire acondicionado | (1) loop motor: 2 1/2 caballos de fuerza
(2) calentador, evaporator, pressure balance mouth, adjustable wind deflector, sensor de temperatura. |
|
| Test cases of standard configuration
|
(1) plane embedded handle,
(2) the door hinge: SUS # 304 import hinge (3) SUS # 304 stainless steel grille type 2 slices steel shelf, single biggest bearing 10 kg. (4) stainless steel SUS # 304 can adjust the spacing between 40 mm shelf track 2 grupos (5) lead hole: phi 50 mm one |
|
| Puerta
|
(1) single door, to open outward, hinges on the left, put his hand on the right (in the face of cabinet positive).
(2) the window/door frame prevent condensation heating device. |
|
| The control panel | 7 inch controller display set and overtemperature protection device, USB and RS232 communication interface | |
| Machine room | Unidades de refrigeración, compressor water pans, exhaust pressure device | |
| Power distribution control cabinet | Distribution board, cooling fan, the total power switch device | |
| The heater | (1) the fin type radiator pipe shaped stainless steel electric heater
(2) the heating control mode: RSS (solid state relay) non-contact cycle such as pulse width (3) the low temperature district heating power: 3.0 kilovatios (4) high temperature heating power: 6.0 kilovatios |
|
| The power cord and drain holes | located on the back of housing | |