Determination of platinum content in platinum ruthenium alloy by energy dispersive X-ray fluorescence spectrometer EDX6000 gold detector
EDX-6000 Precious Metal Jewelry detector Energy Dispersion X-ray Fluorescence spectrometer Gold detector Features:
Small spot (1mm), accurate fixed point test
Vertical light path design, effectively reduce the impact of non-planar samples on test results
Accurate, fast and non-destructive> — Key test, quickly get started
No need to repeat calibration every day, improve the detection efficiency
Expand the application — plating and plating solution analysis
Strong and durable, suitable for complex working environment
Determination of platinum content in platinum ruthenium alloy by energy dispersive X-ray fluorescence spectrometer EDX6000 gold detector
The platinum content in Pt ruthenium alloy (85%~95% PT) was determined by the EDX-6000 gold meter of energy dispersive x ray fluorescence spectrometer. A set of working standard samples with platinum ruthenium as the main content were used. Measure and collect the characteristics of the elements contained in the working standard sample x ray spectrum to establish the standard working curve, compile and optimize the analysis program, determine the content of platinum elements in the platinum ruthenium alloy sample test results compared with other methods, prove the method is feasible. When the platinum content in platinum ruthenium alloy is 85%~95%, the analysis accuracy is better than 0.25%1