XRF fluorescence spectrometer test conditions set guidelines

XRF fluorescence spectrometer test conditions set guidelines

Guidelines for configuring XRF fluorescence spectrometer filters and optical tube pressure flow

XRF spectrometers use Rh X-ray sources and Pd slits. This means that Rh and Pd lines will always be present in your spectrum unless they are removed. The Rh and Pd L lines will mask the presence of elements such as CI and S. Filters can be used to remove Rh and PL lines from the spectrum. Combining light tube power and different filters can also make the instrument more sensitive to certain elements. Recommended power/filter Settings

General analysis of different applications is discussed below

This setting allows X-rays from 1 Zu 40 keV to reach the sample. All elements from Mg to Pu will be excited, producing a fluorescent signal. Rh and Pd lines from the source will appear in the spectrum. This prevents the C and S lines from being observed.Set values:

· Do not use filters

.· Set the voltage to 40 kV

· Use the lowest possible current for metal samples and the highest possible current for non-metal samples · Use a vacuum

Analysis of light elements (Cu and below, except S and CI)

This setting allows X-rays from 1 Zu 15 keV to reach the sample. Mg, Al, Und, and P are best excited for elements of Cu. The Rh and Pd lines will be displayed, so you will not be able to observe the CI or S lines. Because the Rh L lines arrive at the sample, elements with absorption edges below 2.3 keV will be strongly excited, this includes Mg, Al, and SiSet point:

XRF fluorescence spectrometer test conditions set guidelines

● Do not use filters

· Set the voltage to 15 kV

· Set the current to 55 μA

· Use vacuum

Light element analysis (Fe and below, except Ti and Sc, but including S and CI)

Heavy metal analysis (higher Z elements, z.B. Hg, Pb, Br,Als)This setting allows X-rays from 14 Zu 40 keV to reach the sample. This setting is most sensitive to As and Pb. There is little sensitivity to elements below Ca.

Set values:

· Use of filters (titanium and aluminum)

· Set the voltage to 40 kV

Start with 10 μA of current;

If the count rate is too high, lower it do not use vacuum

Analysis of higher Z elements (Fe to Mo) in ceramics

XRF fluorescence spectrometer test conditions set guidelines

This setting allows X-rays from 17 Zu 40 keV to reach the sample. This will excite elements ranging from iron to molybdenum.

The setup is optimized for Rb, Sr, Y, Zr, and Nb. There is little sensitivity to elements below Fe.

Set values:

· Use filter (Cu,Von, Al)

· Set the voltage to 40 kV

· Set the current to 15 μA

· Do not use a vacuum