KDGF-C UHF-Teilentladungstester
- Beschreibung
- Anfrage
Beschreibung
Partial Ddischarge Tester
Produkteinführung
The KDGF-C UHF Partial Discharge tester is a specially developed and simple instrument for detecting electromagnetic wave and ultrasonic activity. Partial discharge An electrical discharge that does not completely short circuit the electrode. The amplitude of this discharge is usually very small. But they do lead to a continuous decline in the performance of the insulation layer, which ultimately leads to the failure of electrical equipment. Non-invasive partial discharge detection provides a fast and easy way to identify potential insulation failures that could cause power outages or cause injury to people. The most practical techniques for non-invasive detection are based on detecting the high frequency part of the electromagnetic spectrum as well as ultrasonic signals. The KDGF-C partial discharge tester is a specially developed and simple instrument for detecting electromagnetic wave and ultrasonic activity.
| Anwendungsbereich | non-invasive detection method is used to detect and locate partial discharge defects of high voltage electrical equipment. |
| detection principle | ultra high frequency method (UHF), ultrasonic method (UA) and ground wave method (TEV). |
| detection frequency band | UHF 300 ~ 1500(MHz), ultrasonic 20 ~ 200(KHz). |
| Messbereich | ultra-high frequency is -80 ~ -20dBm, ultrasonic is 0 ~ 90dB |
| sensitivity | minimum 10pC(depending on the distance between the sensor and the power supply) |
| Sensor | ① Ultra-high frequency sensor: 300 ~ 2000(MHz), with directional reception characteristics; ② Ultrasonic sensor: 20 ~ 200(kHz); (3) Ground wave (TEV) : 10 ~ 70MHz. |
| with built-in ultrasonic sensor, ground wave, ultrasonic two-in-one sensor; | |
| Software functions | (1) Continuously detect UHF, geoelectric and ultrasonic signals to determine whether partial discharge exists; 2.The change trend of the measured signal is displayed in real time, which can make a more intuitive judgment on the development of partial discharge signal; ③ With the on-site data storage function. |
| Eigenschaften des Instruments | ① Screen display: high contrast 3.5-inch TFT color screen. 2.Datenspeicher: 1000 groups of test data can be saved. ③ Working power supply: built-in 8.4V lithium battery, kontinuierlich arbeiten kann 8 Std.. Stromversorgung: Input 100-240VAC, output 8.4V/3A, charging time 3 ~ 4 Std.. ⑤ Dimensions: 220 * 100 * 40. ⑥ Instrument weight: 1.5kg. ⑦ Ambient temperature: -20℃ ~ 45℃. ⑧ Storage temperature: -25℃ ~ 60℃. |
| Komplette Konfiguration | host, Sensor, AC adapter, connecting cable and transport box. |