Beschreibung

Transformer Test System

Kurze Einleitung

GA200XB is professional transformer comprehensive tester. Can be in 50 feet within the electronic transformers, power transformers, network filters, such as all kinds of transformer performance testing. Using 7 inches LCD TFT display. It is a high-speed, broadband, 6-bit test resolution of the impedance measuring instrument, multi frequency points and 0.05% accuracy can meet the requirements of various occasions’ component parameters detection. It is a high cost of quality testing equipment.

Transformer Test System
Technische Parameter
SPEZIFIKATIONEN

GA200XB/GA300XB/GA500XB Transformer Tester

Parameter | Z|, |Y|, C, L, X, B, R, G, D, Q, θ, DCR, TRUN, TURN ratio, Lk, Phase
Basic Genauigkeit 0.05%; DCR, TURN, TURN ratio: 0.1%
Frequenz GA200XB:20Hz~200kHz; GA300XB:20Hz~300kHz; GA500XB:20Hz~500kHz
Anzeigebereich Z,X,R:0.01mΩ~99.9999MΩ DCR:0.01mΩ~99.9999MΩY, G,B:0.00001µS~99.9999S

C:0.00001pF~9.9999F L:0.00001µH~9999.99H D:0.00001~9.9999 Turn ratio:1:0.001-1000:1

Q:0.00001~9999.9 q(rad): -3.14159 - - 3.14159 q(deg): -179.999° -179.999°

Source Resistance 30 Oh, 100 Oh, 10 /100, 10 /CC
Reichweite Auto, Hold, Manual selection
Signal Level 5 mV-2V , Genauigkeit:10%; Constant level:10 mV-1 V, Genauigkeit:5%; TURN: 5 mV-10V , Genauigkeit:10%
DC Bias Source 0V,1.5V,2V,Genauigkeit:1%; Option: CKT100mA:±10V(±100mA)
Triggermodus INT, MAN, EXT, BUS
Geschwindigkeit 75t/s, 12t/s, 3 t/s
Equivalent Circuit In series and parallel
Result Display Direct,Δ,Δ%,V/I
Kalibrierung Open / short / Belastung
Schnittstelle RS232C, USB-HOST, USB DEVICE, Headset interface, Foot switch; Optional: HANDLER,GPIB,LAN interface
Andere The software version of the instrument can be upgraded and updated by U disk.

Maximum can measure transformer products with 50 pins

ALLGEMEINE MERKMALE
Stromversorgung Stromspannung: 110V/220V; Frequenz: 50Hz – 60Hz

375mm (W) *135mm (H) *350mm (L); 7.5kg

Size and Weight
BESTELLINFORMATIONEN
Zubehör 20PIN transformer scanning box (Handbuch, automatic two optional); Transformer scan control line;

Four terminal test cable; Foot switch; AC Power Cord