Technische Spezifikation

Test station 2group or 2 layer
Test stroke 0-500mm
Testgeschwindigkeit 5-10times/min
Timer 0-999999999 Power off memory protection function, set the number of shutdown function
Action principle Horizontal drawer movement
Leistung AC220V 50HZ
Air pressure 6kg
Out-Größe 1200*500*700mm
Gewicht 60KG

GA-9001-A Digitaler Sauerstoffindextester

Scope of application:

It is suitable for the determination of the combustion performance of plastics, Gummi, Faser, foam plastics, film and textile materials

GA-9001-A Digitaler Sauerstoffindextester

Wichtigste technische Parameter

  1. Using imported oxygen sensor, digital display oxygen concentration without calculation, more accurate and more accurate, range 0-100%
  2. The built-in temperature controller to control the temperature, the control range: 23 ℃ ±2 ℃
  3. Digital resolution: 0.1%
  4. Meßgenauigkeit: 0.4 Grad
  5. Flow regulation range: 0-10l /min (60-600l /h)
  6. Response time: < 5S
  7. High temperature resistant quartz glass tube, inner diameter 75 ㎜ height 480 mm or more
  8. Gas flow rate in combustion cylinder: 40mm ±2mm/s
  9. Pressure gauge accuracy 2.5, Auflösung: 0.01MPA
  10. The flowmeter: 1-15 l/min (60-900 l/H) ist einstellbar, test environment: the environment temperature: room temperature ~ 40 ℃;, relative Luftfeuchtigkeit: 70%;
  11. Input pressure: 0.2-0.3MPA
  12. Working pressure: nitrogen 0.1-0.15mpa oxygen 0.1-0.15mpa oxygen/nitrogen mixed gas inlet: including pressure regulator valve, flow control valve, gas filter and mixing chamber.
  13. The sample holder can be used for soft and hard plastics, textiles, fire doors, usw
  14. Propane (butane) ignition system, flame length 5mm-60mm can be adjusted freely
  15. Gas: nitrogen and oxygen for industrial use with purity of > 99%;(user supplied).
  16. Power requirements: AC220 (±10%) V, 50HZ
  17. Maximum working power: 50W
  18. Igniter: made of a metal, end diameter Φ 2 ±1 mm nozzle, can ignite sample in combustion tube, the flame length: 16 ±4 mm, the size is adjustable
  19. Self-supporting material sample holder: it can be fixed in the position of the axis of combustion cylinder and can hold the sample vertically
  20. Non-self-supporting material sample holder: two vertical edges of the sample can be fixed on the frame at the same time (optional for non-self-supporting materials such as textiles and films).