Beschreibung

Flash Point Tester

Produkteinführung

GA-6054-B automatic opening flash point analyzer is used to determine the opening flash

point value of petroleum products.The instrument adopts electronic ignition mode and does

not require any combustible gas, which fully conforms to the standard method requirements

of GB/T3536-2008 and GB/T267-88.The instrument adopts high-performance ARM

microprocessor and fuzzy PID technology for temperature control. The temperature rising

speed is uniform and stable. The instrument also adopts touch color LCD, which makes

human-computer interaction more convenient.It is a special instrument for petroleum, elektrisch

Leistung, Chemieindustrie, commodity inspection and other industries to replace imported

Produkte.

Flash Point Tester

Leistungsmerkmale

1.Hoher Automatisierungsgrad: automatic heating, sweeping, lifting, Kühlung, printing,

calibration test results;

2.Genauigkeit der Temperaturregelung: adopt advanced fuzzy PID temperature control technology;

3.Adopt advanced ion ring flashover detection technology;

4.Test sample opening flash point value;

5.Test overtemperature self-stop;

6.Instrument self-test function;

7.Luminance adjustment function of igniter;

8.Große Datenspeicherkapazität, 3000 analysis and test records can be saved;

9.Data record printing function.

Flash Point Tester

Technische Indikatoren

1.Standard: GB/T3536-2008 GB/T267-88

2.Detection item: opening flash point

3.Anzeigemodus: touch high – definition color LCD

4.Messbereich: room temperature ~ 400℃

5.Auflösung: 0.1℃

6.Wiederholbarkeit: ±3℃

7.Hebemodus: pitch type

8.Ignition mode: electronic ignition (without gas)

9.Data records: 3000

10.Printing: Thermal printer

11.Applicable ambient temperature: 5 ~ 35℃

12.Applicable relative humidity: ≤85%

13.Energiequelle: AC220V±10%, 50Hz±5%

Work rate: 550W

14.External dimension: 435*350*280mm

15.Gewicht: 15Kg

Flash Point Tester