Beschreibung

Ozone Climatic Chamber

Technisch PParameter

1.Product Name Ozone Climatic Chamber
2.Modell GA-2013-B
3.Limit sample The test equipment is prohibited:

(1) Test or storage of samples of inflammable, explosive and volatile substances

(2) Test or storage of corrosive substances

(3) Testing or storage of biological samples

(4).The test or storage of strong electromagnetic emission source samples

 

Ozone Climatic Chamber

4.Volumen, Abmessungen , Gewicht

4.1.Nominal content product 150L
4.2.The type size W500*H600*D500mm
4.3.External dimension W950*H1750*D1000mm
4.4.Gewicht

 

About 250KG
Ozone Climatic Chamber

5.Performance

 

5.1Test Environmental Conditions

 

5.2.Test Method

The ambient temperature is +25℃, relative humidity ≤85% and there is no sample in the test chamber

* The stainless steel test chamber meets the international standard and the completely closed gas loop for producing ozone by silent discharge. The entire ozone gas generation and capture is a closed loop to ensure personal safety.

* arbitrarily set ozone concentration in the range of 1000pphm.

* Sample holder device driven by speed regulating motor

(It satisfies the static tensile test, and the specimen’s running trajectory satisfies the cycloidal trajectory principle)

* Ozone concentration automatic digital display control instrument.

* Double beam “QUA” non-dispersive ultraviolet ozone detector.

* Digital display automatic online measurement.

* There is a calibration interface for metrological verification.

* Ozone concentration/temperature and humidity setting value control PID self-setting, convenient and reliable operation.

* 3rd generation Bill Lambert measurement application software.

* Security door lock control system

 

5.3.Ozone concentration range 20~500 pphm(Set up direct reading)
5.4.Ozone concentration deviation Ozone concentration deviation: national standard 10%
5.5.Test chamber temperature Messbereich: room temperature ~ +70℃ (temperature range when ozone occurs)
5.6.Relative Luftfeuchtigkeit  

Measurement range below 60%RH (humidity range when ozone occurs)

 

5.7.Specimen holder 360 degree rotating sample stand (static)
5.8.Dynamic stretching Each fixture can install test pieces of tensile type can meet Ⅰ dumbbell dynamic work piece and the actual workpiece
5.9.Test gas flow rate 0-60L/Min
5.10.Temperaturbereich 0℃→+100℃(Arbitrary setting)
5.11.Humidity range 20-65%RH
5.12.Temperaturschwankungen ±0.5℃
5.13.Humidity fluctuation ±2.5%
5.14.Temperature Departure ±2.0℃
5.15.Humidity Deviation ±3.0%RH
5.16.Condition of Loading Not have
5.17.Working noise A levels of 70 db (A) oder weniger

(Measured in a sound-proof room with ambient temperature of 25℃ and little echo;A was used to calculate weight and test the average value of 8 points.Each test point is 1 m horizontally away from the noise source and 1 m above the ground.)

5.18.Meet test standard Equipment used for testing ozone aging performance of rubber and its products:JIS K 6259 ASTM1149 ISO1431 GB/T7762 GB/T13642

Ozone Climatic Chamber

6.Architectural Feature:

6.1.Insulation enclosure:(1) Outer wall material: A3 steel plate with electrostatic spraying

(2).Inner wall material: mirror stainless steel plate SUS# 304.

(3).Box insulation material: 100mm high temperature resistant rigid polyurethane foam.

(4).Insulation material for box door: 100mm high temperature resistant rigid polyurethane foam.

6.2.Bottom structural strength:The heavy capacity of track at the bottom of the test box: ≤100Kg/m (Belastung). The bottom plate of the studio is welded with 4mm channel steel.

6.3. Air conditioning passage Stainless steel long axis centrifugal fan: 1 /90W.

Fan, heater, evaporator (and dehumidifier), drainage device, pressure balance port, adjustable air guide plate, Temperatursensor

6.4.(1).Standard configuration of test box Window 250x350x40mm 3 layers of vacuum toughened glass.

(2).Flush type handle,

(3).Door hinge :SUS #304 inlet hinge

(4).Energy-saving light in the box: LED light emitting way

(5).Lead hole: 50mm 1 (with a rubber plug)

(6).Ozone discharge outlet: 50mm towards outdoor

Ozone Climatic Chamber Ozone Climatic Chamber Ozone Climatic Chamber