STD-A Dielectric loss Meter
- Popis
- Poptávka
Popis
STD-A Dielectric loss Meter
Úvod
Dielectric loss and dielectric constant are important physical properties of various ceramic such as electro ceramics,ceramic for mounting purposes,capacitor etc. The instrument is used in determining dielectric loss angle tangent tan δ and dielectric constant (E) to know more about various factors that effect the dielectric loss and dielectric constant, and to provide basis for improving the properties of materials. It is widely used in application study of properties of inorganic nonmetal new materials of institutions of higher education and research institutions. The instrument is in line with the standard of GB/T5594.4-1985.
STD-A Dielectric loss Meter
Main Technical Index:
1. The Q value measurement
A. Q value measuring range: 5 ~ 999.
B. Q value range step: 30, 100, 300, 999, automatic shift.
C. the nominal error
Frekvenční rozsah: 25 kHz ~ 10 MHz;
Inherent error: ≤5%±the full value of 2%;Work of error: ≤7%±the full value of 2%;
Frekvenční rozsah: 10 MHz ~ 50 MHz;
Inherent error: ≤7%±the full value of 2%;Work of error: ≤10%±the full value of 2%;
2. The inductance measurement
A. rozsah měření: 0.1 mu H ~ 1 H.
B. grade divide: seven ranges.
0.1 ~ 1 u H, 1 ~ 10 u H, 10 ~ 100 u H,
0.1 ~ 1mH, 1 ~ 10 mH, 10 ~ 100 mH, 100mH ~ 1 H.
3. The capacitance measurement
A. rozsah měření: 1 ~ 460 pF (výše 460 pF capacitance measurement see using rules);
B. the adjusting range capacitance
Main tuning capacitor: 40 ~ 500 pF;
Přesnost: níže 150 pF ±1.5pF; výše 150 pF ±1%;
Fine-tuning capacitance: – 3 pF~ 0 ~ + 3 pF;
Přesnost: ± 0.2 pF.
4. The oscillation frequency
A. oscillation frequency range: 25 kHz ~ 50 MHz;
B. the frequency step:
10 ~ 99 kHz; 100 ~ 999 kHz;1.0 ~ 9.99 MHz;10 ~ 50 MHz.
C. frequency error: 2 X 10-4 ± 1 digit.
5. Q qualified preset functions indicated, preset range: 5 ~ 999.
6. Equipment normal working conditions:
A) teplota: 0 ℃ ~ + 40 ℃;
B) relativní vlhkost: 80%;
C) zdroj napájení: 220 V±22V, 50 Hz ±2.5 Hz