Popis

Optical Emission Spectrometer
Application Field

LR-8681 is a full-spectrum direct-reading inductively coupled plasma emission spectrometer for the determination of trace and trace elements in different substances (dissolved in nitric acid, hydrochloric acid, hydrofluoric acid, atd.). It is widely used in environmental protection, petroleum products, rare earth, semiconductor, geology, metallurgy, chemical industry, clinical medicine, jídlo, biological samples, criminal science, agricultural research and other fields.

Výkonové charakteristiky
Performance stability
The use of all-solid-state rf power supply, s malou velikostí, vysoká účinnost, stable output power, with a variety of protection
functions and many other advantages, the load using fully automatic matching technology, matching speed, improve the efficiency of the use of power supply and the stability of the instrument, and make the whole ignition process simple and convenient.

Injection automation
Adopting four-channel automatic design, the rotational speed can be set up according to the test demand to adjust the flow rate, carrier gas, plasma gas, auxiliary gas are all controlled by advanced mass flow controller, at the same time can be equipped with imported high salt nebulizer, imported hydrofluoric nebulizer and so on to meet the customer’s various test needs.
Precision analysis
The middle step grating prism cross dispersion mode is adopted, there is no moving optical element, the ultra low stray light design is combined with the unique optical design, nitrogen distributed blowing, imported optical elements, intelligent and accurate automatic wavelength calibration algorithm.
Optical Emission Spectrometer
RFPower technical parameter
circuit type solid-state RF Power supply, with function of automatch
Frekvence 27.12MHz±0.05%
Frequency Stability <0.1%
Power Output: 800W-1500W
Power Output Stability <0.3%
Escaped RF radiation 30cm away from the instrument, electric field: E<2V/m

Sampling System Technical Parameter

Output working coil inner diameter 25mm
Torpue tube Three concentric, external diameter 20mm
Coaxial nebulizer Outer diameter 6mm
Double barrel atomizing chamber Outer diameter 34mm

Optical Emission Spectrometer Emission Spectrometer

Gas Flow Controls
Plasma Argon Flowmeter (100-1000)L/h(1.6-16L/min)
Auxiliary Argon Flowmeter (100-1000)L/h(0.16-1.6L/min)
Auxiliary Carrier Argon Flowmeter (100-1000)L/h(0.16-1.6L/min)
Pressure Maintaining Value (0-0.4MPa)
Cooling Water Teplota:Teplota:20-25℃,Rate of Flow>5L/min,Hydraulic Pressure>0.1MPa
Technical index of spectrometer
Grating Middle step grating,52.67 lp/mm,64 sparkle angle
Rozsah vlnových délek 160-1000 nm.
Numerical aperture F ≤ 8, ultra-high luminous flux to ensure the detection limit and sensitivity of the instrument
Rozlišení < 0.0065nm@200nm
Astigmatism Equivalent background concentration of 10000 ppm Ca solution at As189.042 nm <2 ppm●Light chamber:Precision constant
teplota,35±0,1 °C,Distributed nitrogen purging, normal purging 1.8L/min, fast purging 3.8L/min
Testing device technical specifications
Target Size 27.6mm ×27.6mm,1024×1024 addressing detection units
Detektor CID
Reading mode Non-destructive read (NDRO), full reading (FF) and arbitrary read integral (RAI)

Optical Emission Spectrometer

Standard Configuration List

Standard parts
Ne Spare parts qty unit notes
1 spectrophotometer      
1.1 ICP8681 1 soubor Line grating 3600L/mm,2400L/mm
2 software      
2.1 [ICP analysis software]V1.0 1 soubor CD with latest version
2.2 Záruční list 1 soubor Angličtina
2.3 Quality card 1 soubor Angličtina
2.4 USB Flash Drive 1 kus  
3 accessoris      
3.1 Cable 1 čára size 8Pin X L3M,RJ-45
3.2 Plastic drum 1 kus Volume 5L plastic:HDPE semi-transparency,white color with white cover,deltoid
3.3 Argon pressure reducing valve 2 kus v:0~25MPa,Out:0~0.4MPa innerG5/8 outter M16×1.5RHM152IN-40
3.4 PU tube 30 meter 10×6.5×1.75mm
3.5 PU tube 5 meter 6×4×1mm
3.6 PVC transparent tube 1 meter 8mm×6mm
3.7 plug 2 kus 10A250V
3.8 Ground copper braided wire 15 meter sectional area:4mm2
3.9 Brass tube 1 kus Size 48,thin 3,L:500
3.10 Filter element of filter tank 4 kus outter:60mm, inner:28mm, L:125mm
3.11 Torch pipe 2 kus  
3.12 Double cylinder atomizing chamber 2 kus  
3.13 atomizer 1 kus  
3.14 Capillary tube 0.5 meter D=2.38mm,d=0.8mm
3.15 Pump head hose of Peristaltic Pump 3 čára Ф1.65 L45cm tygon-R-3607(2S)
3.16 Electronic Fuses 2 čára 10A,250PROTI,6×30mm
3.17 Electronic Fuses 2 čára 6.3A,250PROTI,5×20mm
3.18 Cylindric atomizer 2 kus PTFE
3.19 Silicone tube 0.3 meter Inner 2mm,outer 4mm,tvrdost:55HRC
3.20 Hose adapter 4 kus ZCJH-2*6 INNER 6-2
3.21 Hose adapter 4 kus ZC-1/16*3/32 INNER 1.6-2.4
3.22 Spherical connector clamp 1 P L00091181,Ф=10
3.23 O shape seal ring 2 P 32.5×3.55
3.24 O shape seal ring 4 P 5.6×1.8
3.25 Quick Connector 2 P Φ6
3.26 Raw material belt 1 roll 26mm×20m,thin:0.1~0.2mm
3.27 ribbon 30 čára Size 4X200
3.28 toolkit 1 soubor Self-made
4 Standard instrument      
4.1 počítač 1 soubor Lenovo
4.2 displayer 1 soubor Lenovo
4.3 Canon Printer 1 soubor Laser printer
4.4 cooling water tank 1 soubor refrigerating capacity:2000W
4.5 stabilized voltage supply 1 soubor 10kVA
5 Packing material      
5.1 ICP6800 wood carton 1 soubor  
5.2 screw 5 P M5×8

Optical Emission Spectrometer Optical Emission Spectrometer Optical Emission Spectrometer