Popis

Two Box Thermal Shock Chamber

Specifikace

Název produktu Two van hot and cold impact testing machine

Use of low temperature and high temperature hot and cold storage tank, in accordance with the cylinder, the hanging basket in turn to high temperature and low temperature box, so as to achieve rapid temperature shock effect, balance (BTC) + special temperature control system design of supply air circulation system, to control SSR P.I.D. way, make the system of heat equal to the amount of heat loss, thus the use of long-term stability.

Model produktu TSD – 11 – A – 2 p (air-cooled type)

Poznámka: air-cooled type need not install cooling towers, water cooled type to install, depends on customer’s requirement.

Limit sample The test equipment prohibited:

(1)Which is inflammable, explosive, easy to volatile substances sample test or store

(2) the corrosive substance sample test or store

(3) biological sample test or store

(4) strong electromagnetic emission source sample test or store

Two Box Thermal Shock Chamber

Hlasitost, size and weight

Nominal content product 52L
Shopping basket size W50*H32*D32cm
Installation dimensions W132*H187*D155CM
hmotnost About 400㎏
Two Box Thermal Shock Chamber

Výkon

Test environment conditions Environment temperature is 25 ℃, relative humidity of 85% or less, test under the condition of no sample in the box

GB/T 5170.2 1996 temperature testing equipment.

High greenhouse +60℃→+150℃
Time +60℃→+150℃ ≤25min
Teplotní rozsah -55℃~-10℃
Time +20℃ → -55℃≤60min
Test (style) Test chamber upside for high temperature, lower for the low temperature zone, specimen is placed in the basket, basket turns in the high temperature box and low temperature in experiment
Shock temperature range (+60~+100)℃/(-40~-10)℃
Temperature fluctuation degree ±0,5 ℃
The temperature deviation ±2,0 ℃
The temperature recovery time ≤5min
Temperature conversion time ≤10sec
Restore conditions High temperature exposure: + 150 ℃ / min – > low temperature exposure: 30 na 55 ℃ / 30 min

Sample maximum weight: 2.5 KG

Meet the test standard GB/T 2423.1 2008 test: A test method at low temperature

GB/T 2423.2 2008 test B: high temperature experiment method

GB/T2423.22-2002 test N: temperature change test method test Na

GJB 150.5 A – 2009 military equipment laboratory test method for part 5: the temperature impact test

GJB 360 A – 1996 temperature impact test

GJB367.2-2001 military communication equipment general specification temperature impact test

IEC 68-2-14 test method for N: temperature changes

Two Box Thermal Shock Chamber

Strukturální vlastnosti

Thermal insulation enclosure structure (1) the outer wall material: advanced electrolytic plate surface treatment of the lacquer that bake.

(2) the lining material: stainless steel plate SUS ﹟ 304.

(3) low greenhouse body heat preservation material, high temperature resistant rigid polyurethane foam.

(4) high greenhouse body heat preservation material: glass fiber high temperature resistant white cotton.

(5). The door is thermal insulation material: glass fiber

Air conditioning channel (1) loop motor: 2 1/2 HP

(2) heater, evaporator, pressure balance mouth, adjustable wind deflector, temperature sensor.

Test cases of standard configuration

 

 

(1) plane embedded handle,

(2) the door hinge: SUS # 304 import hinge

(3) SUS # 304 stainless steel grille type 2 slices steel shelf, single biggest bearing 10 kg.

(4) stainless steel SUS # 304 can adjust the spacing between 40 mm shelf track 2 skupiny

(5) lead hole: phi 50 mm one

Door

 

(1) single door, to open outward, hinges on the left, put his hand on the right (in the face of cabinet positive).

(2) the window/door frame prevent condensation heating device.

The control panel 7 inch controller display set and overtemperature protection device, USB and RS232 communication interface
Machine room Refrigeration units, compressor water pans, exhaust pressure device
Power distribution control cabinet Distribution board, cooling fan, the total power switch device
The heater (1) the fin type radiator pipe shaped stainless steel electric heater

(2) the heating control mode: SSR (solid state relay) non-contact cycle such as pulse width

(3) the low temperature district heating power: 3.0 KW

(4) high temperature heating power: 6.0 KW

The power cord and drain holes located on the back of housing