Popis

Měřič kontaktního úhlu GA-1101

souhrn

Contact Angle is the tangent line of the gas-liquid interface at the intersection point of gas, kapalné a pevné. This tangent line has an Angle between the liquid and the solid-liquid boundary. The measurement of contact Angle is the main method for surface performance detection nowadays.

GA-1101 basic contact Angle measuring instrument adopts the principle of optical imaging to measure the surface contact Angle, smáčivost, surface energy and other properties of the sample by means of image profile analysis. The equipment has high cost performance, comprehensive functions and can meet the requirements of conventional sample surface wettability measurement.

 

Měřič kontaktního úhlu GA-1101

aplikace

Contact Angle measuring instrument is widely used in various industries, in the mobile phone manufacturing, výroba skla, povrchová úprava, materiálový výzkum, chemical and chemical industry, výroba polovodičů, paint and ink, electronic circuits, textilní vlákno, medical biological and other fields, Měření kontaktního úhlu se stalo důležitým nástrojem pro hodnocení vlastností povrchu.

  1. Wetting behaviors of liquid on solid surface, jako je šíření, infiltration and absorption, atd., static contact Angle was measured by droplet method;
  2. Measurement of forward Angle, zpětný úhel, Kontakt Úhel zpoždění, rolling Angle and dynamic contact Angle of materials on solid surface;
  3. Continuous real-time research and process recording of absorption materials, and curve analysis of contact Angle changes with time;
  4. Measurement of contact Angle of various special materials, jako je prášek, zakřivený povrch, superhydrophobic/superhydrophilic samples;
  5. Test the contact Angle of the material immersed in the liquid by adhesive drop method;
  6. Measurement of interfacial tension, polarity and dispersion components of various liquid surface by droplet method;
  7. Calculate the Surface free energy and analyze its polar dispersion component;
  8. Analyze the adherent work of liquid on solid surface to evaluate the uniformity and cleanliness of solid surface.

 

Měřič kontaktního úhlu GA-1101

Specifikace zařízení

specification of equipment
Device host
Boundary dimension 420mm (L) *150mm (W) *400mm (H)
Hmotnost 3.2KG
Power supply
Napětí 100~240VAC
Napájení 20W
frekvence 50/60HZ
nterspace
platform size 130mm × 150 mm
Maximum sample 180mm×∞×30mm
Sample table regulation 3D manual adjustment (upgradable automatic)
Ruční nastavení, zdvih 60 mm, accuracy 0.1mm

Left and right manual adjustment, zdvih 35 mm, accuracy 0.1mm

Manual up and down adjustment, zdvih 80 mm, accuracy 0.1mm

Picture pick-up system
Image Maximum 3000(H)× 2000(V)
Maximum frame rate 70fps
senzor SONY 1/1.8″
spektrum Black and white/Color
ROI custom
Display line weight custom
time of exposure custom
zdroj napájení 5 VDC USB port
transfer USB3 Vision
Microscope objective
focal length 100mm
multiplying power 8X
Resolution scale 6~12um
Light source
typ Single wavelength Industrial LED (Cold light)
wave length 460nm
light field 40mm×20mm
lifetime 50000Hour
Lifetime
Drip system Direct injection precision microinjector
režim ovládání manual control
Drop accuracy 0.1μl
injector High precision airtight syringe
kapacita 1000μl
syringe needle 0.51mm stainless steel super hydrophobic needle (Standard)
Software
Contact Angle range 0~180°
poměr rozlišení 0.01°
Contact Angle measurement method Fully automatic, semi-automatic, manuál
analytický režim Stop drop method (2/3 stát), bubble trapping method, seat drop method
analytical method Static analysis, dynamic analysis of increasing and shrinking liquid, dynamic analysis of wetting, real-time analysis, bilateral analysis, forward and backward Angle analysis
test method Circle method, ellipse/oblique ellipse method, differential circle/differential ellipse method, Young-lapalace, width and height method, tangent method, interval method
Surface free energy
test method Zisman, OWRK, WU, WU 2, Fowkes, Antonov, Berthelot, EOS, adhesion work, wetting work, spreading coefficient
Data processing
output mode Automatically generated, can export/print EXCEL, Word, spectrum and other report formats