Описание

LT3600 лазерен тестер за размер на частиците

Patented technology:
◆ Polarization filtering technology *
◆ Compensation and correction technology of anomalous change of diffraction Airy spot (ACAD)
◆ Inclined trapezoidal measuring window *
◆ Grille type large Angle detection technology
◆ Particle size analysis model optimization and adaptive technology
◆ Dual drive injection decentralized integration technology

 

LT3600 лазерен тестер за размер на частиците

Performance :
◆ Fully in line with ISO13320 diffraction measurement technical standards
◆ Unique optical path configuration, large continuous physical measurement Angle, no detection blind area
◆ Improved inversion algorithm, users do not need to choose “режим на анализ”, taking into account extremely high resolution and stability
◆ No need to replace the lens, no need to use standard sample calibration, measuring range up to 0.02 micron to 3600 micron
◆ Holographic signal synchronous processing technology, real-time measurement speed of up to 20,000 times per second, does not miss any shape and distribution of particles diffraction information
◆ The ultra-stable solid laser light source system with automatic temperature constant technology completely overcomes the disadvantages of long preheating time and short service life of He-ne gas laser
◆ The polarization space filter technology is used to completely abandon the pinhole filter that leads to poor mechanical and thermal stability
◆ High strength metal optical base and laser energy tracking and stabilization technology, with high stability and anti-interference ability

 

LT3600 лазерен тестер за размер на частиците

Параметър

принцип на измерване laser diffraction
optical model Full range Mistral theory and Fraunhofer theory are optional
particle size range 0.02μm-3600μm, no need to replace the lens, does not rely on standard sample calibration
detecting system Contains a grille type large Angle, non-uniform cross area compensation detector array, the whole measurement Angle range without blind area
метод на вземане на проби ** Wet dispersed sampling system: Support SOP one-click operation measurement, compatible with organic and aqueous dispersion media, sample pool capacity of up to 1000 мл, liquid level continuously adjustable. Sample mixing and circulation dual motor design and independent control, speed 0-2500RPM continuously adjustable. Built-in high-power probe type ultrasonic dispersion system, the dispersion power is continuously adjustable, and supports automatic bubble elimination technology. (Optional Hydrolink injection system)
** Dry dispersed sampling system: Support SOP one-click operation measurement, dispersion pressure 0-4.5Bar continuous adjustable, airflow pressure control accuracy ±0.1Bar (optional Aerolink injection system)
източник на светлина 638nm solid-state laser with integrated thermostatic system, up to 20mW
Spatial filtering mode Non-pinhole polarization filtering technology
Optical alignment system Intelligent fully automatic
measuring time Typical values are less than 10 секунди
measuring speed Up to 20,000 times per second
degree of accuracy Dv50 better than ±0.6% (NIST-Traceable latex standard)
повторяемост Dv50 better than ±0.5% (NIST-Traceable latex standard)
laser safety Клас 1 laser products
computer configuration Intel i5 processor, 4GB RAM, 250GB hard drive, мишка, keyboard and widescreen display
computer interface USB2.0 or above
Software running platform Windows7 or above Professional Edition
operating ambient temperature 5℃ -40 ℃
Работна влажност на околната среда 10%-85% relative humidity (без конденз)
power requirement Ac 220V, 50Hz-60Hz, standard ground
Optical system weight 28килограма
Optical system size 636mm x 275mm x 320mm