GA-2001-D Programmable Temperature Humidity Chamber 80L
- Описание
- Разследване
Описание
GA-2001-D Programmable Temperature Humidity Chamber 80L
Design Standard:
| GB11158 | High Temperature Test Chamber Technical Condition |
| GB10589-89 | Low Temperature Test Chamber Technical Condition |
| GB10592-89 | High-low Temperature Test Chamber Technical Condition |
| GB/T10586-89 | Damp Heat Test Chamber Technical Condition |
| GB/T2423.1-2001 | Low Temperature Test Chamber Test Methods |
| GB/T2423.2-2001 | High Temperature Test Chamber Test Methods |
| GB/T2423.3-93 | Damp Heat Test Chamber Test Methods |
| GB/T2423.4-93 | Alternating Damp-Heat Test Methods |
| GB/T2423.22-2001 | Temperature Changing Test Methods |
| IEC60068-2-1.1990 | Low Temperature Test Chamber Test Methods |
| IEC60068-2-2.1974 | High Temperature Test Chamber Test Methods |
| GJB150.3 | High Temperature Test Methods |
| GJB150.4 | Low Temperature Test Methods |
| GJB150.9 | Damp Heat Test Methods |
GA-2001-D Programmable Temperature Humidity Chamber 80L
Uses:
This equipment can simulate different environment condition. It’s appropriate for testing material performance, such as resist heat, resist dry,resist humidity and resist cold. That can define performance of material.
GA-2001-D Programmable Temperature Humidity Chamber 80L
Спецификация
| Модел | GA-2001-D-80L
(A~F) |
GA-2001-D-
150Л (A~F) |
GA-2001-D-
225Л (A~F) |
GA-2001-D-
408Л (A~F) |
GA-2001-D-
800Л (A~F) |
GA-2001-D-
1000Л (A~F) |
| Inner size: WHD(см) | 40*50*40 | 50*60*50 | 60*75*50 | 60*85*80 | 100*100*80 | 100*100*100 |
| External size: WHD(см) | 105*165*98 | 105*175*108 | 115*190*108 | 135*200*115 | 155*215*135 | 155*215*155 |
| Температурен диапазон | (Low temp.:А:+25℃; б:0℃;° С:-20℃; д:-40℃; д:-60℃; Е:-70℃) (High temp.: + 150℃) | |||||
| Humidity range | 20%~98% R.H | |||||
| Temp analytic accuracy/
uniformity |
0.1℃/±2.0℃ | |||||
| Temperature Fluctuation | ±0.5℃ | |||||
| Humidity control accuracy | ±0.1%;±2.5% | |||||
| Heat, cooling time | Heat up about 4.0°C/min ; Cooling about 1.0°C/min | |||||
| Inner and outer materials | SUS#304 Stainless steel for inner chamber ; Carton Advanced cold plate nano paint for outer | |||||
| Insulation materials | Resistant to high temp ,high density,formate chlorine,ethyl acetum foam insulation materials | |||||
| Cooling system | Air cooling/single segment compressor(-40°C) ,air and water double segment compressor
(-50°C~-70°C) |
|||||
| Protection devices | Fuse switch, compressor overload switch, refrigerant high and low pressure protection switch,
super humidity over-temperature protection switch, предпазител, failure warning system |
|||||
| Parts | Watching window , 50mm testing hole , PL internal bulbs , Wet and dry bulb gauze , partition plate , castorx4, Foot Cupx4 | |||||
| Compressor | Original France“Tecumseh”brand | |||||
| Контролер | Тайван, Independent R&D software | |||||
| Мощност | AC220V 50/60Hz & 1 , AC380V 50/60Hz 3 | |||||
| Тегло (килограма) | 170 | 220 | 270 | 320 | 450 | 580 |
| 1. Вещ
2. Модел |
Constant Temperature and humidity test chamber
80Л |
| 3. функция | It simulates the products in the chamber where the climate condition of different temperature and humidity, such as high-low operation and storage, temperature cycling, high-low temperature and high-low humidity, moisture condensation etc. to check its adaptability and changing performance.
It’s required to conform to international regulation (IEC, ТОЙ, GB, MIL…) to reach the consistency (including test schedule, condition, метод) under the international standard. |
| 4. Sample application | The chamber applies to:
compatibility test for store, transportation and application of electric, electronic products, unit components, parts and other materials under high or low temperature. The following samples are prohibited to be tested and stored: – Inflammable , explosive and other volatile substances; – Corrosive substance; – Biont; – Strong electromagnetic emission source. |
| 5. Сила на звука, размер & тегло | |
| 5.1. Сила на звука
5.2. Internal size 5.3. Exterior size 5.4. Тегло |
80Л
У 400 x H500 x D400mm W1050 x H1650 x D980mm 170КИЛОГРАМА |
| 6. Performance | |
| 6.1. Test condition | температура:+25℃, Влажност: ≤85%, no samples in the chamber |
| 6.2. Метод на изпитване | GB/T 5170.2-1996 Temperature test chamber
GB/T5170.5-1996 Temperature and humidity test chamber (only temperature and humidity) |
| 6.3. Температурен диапазон | -40℃→+150℃ (регулируема) |
| 6.4. Температурни колебания | ±0.5℃ |
| 6.5. Temperature accuracy distribution | ±2.0℃ |
| 6.6. Humidity range | 20%~98%R.H (temperature at 20℃-85℃) |
| 6.7. Humid. fluctuation | ±2.5%R.H. |
| 6.8. Humidity accuracy distribution | ±3.0%. |
| 6.9. Heat-up time | -40℃ → +150℃ Heat up for 2-4.0℃/min |
| 6.10. Cooling time | +150℃→ -40℃ Cooling for 0.7-1.0℃/min |
| 6.11. Load condition | Testing space is not more than 3/4 of the total space, no heating |
| 6.12. Sample holder | Stainless steel punching stencil |
| 6.13. Relative humidity and temperature control range (refer to fig.) | |
| 6.14. Meet the technical standards
|
GB/T 2423.1-2001 Test A: Low temperature test method
GB/T 2423.2-2001 Test B: High temperature test method GJB 150.3-1986 High temperature test GJB 150.4-1986 Low temperature test IEC68-2-1 Test A: Cold IEC68-2-2 Test B: Dry heat GB11158 “Technical Conditions for High Temperature Test Chamber ” GB10586-1 “Technical conditions for Damp heat test chamber ” GB/T 2423.2 “Basic environmental testing procedures for electric and electronic products B: High temperature test method”. GB/T 2423.3 “Basic environmental testing procedures for electric and electronic products Ca: Constant damp heat testing method” etc. |
| GA-2001-D Programmable Temperature Humidity Chamber 80L
7.Structure characteristics. |
|
| 7.1. Insulated enclosure structure
|
Външен материал: SUS304 or stoving varnish
Inwall material: SUS304 Insulation material of box body: flame retardant fire rigid polyurethane foam Door thermal insulation material: flame retardant fire rigid polyurethane foam |
| 7.2. Air conditioning channel | Centrifugal fan:
Heater, evaporimeter, outfall, temperature sensor |
| 7.3. Test chamber standard
|
Window 330x450x40mm 3 vacuum layers
Anti-explosion handle Door hinge: SUS #304 Lamps in the chamber: 11У Access port: φ50mm or φ100mm 1pc or more. |
| 7.4. Door
|
Single door, open from left, handle is at right; With viewing window: W300xH450mm, lamp window/condensation prevention thermal device |
| 7.5. Контролер | Control screen, overtemperature protection setting |
| 7.6. machine room | Refrigeration units , water pans , outfall |
| 7.7. control cabinet for power distribution | Switchboard fan, ventilator |
| 7.8. Heater | Heating system: cycling heating from finned-tube Stainless steel electric heater |
| 7.9. Humidifier | Oven humidification system, cycling winding system |
| 7.10. Power line hole and outfall position | located at the back of chamber |