- وصف
- سؤال
وصف
DRX-II-300A Thermal Conductivity Machine
مقدمة
It is mainly used to test the thermal conductivity of construction insulation materials, زجاج, fiber, foam and so on. Testing materials include fiber board, loosing filler glass fiber, ceramic fiber (XPS/EPS/PUR, إلخ.), foam plastic, inter-layer composite plate, vacuum plate, rubber products, cement products, concrete brick, powder materials, filler materials, gypsum board and so on. The tester is in line with the latest national standard GB/T 10294-2008 guarded hot plate apparatus, ASTM C518, ASTM C177 and ISO 8302 Standard Test Method for Steady-State Heat Flux Measurements and Thermal Transmission Properties by Means of the Guarded-Hot-Plate Apparatus. It is in full compliance with the national thermal conductivity measurement standard with its stable and accurate temperature control technology, computer interface, software’s real-time monitoring and control, data processing and saving connecting. With these advantages, it is applied widely in materials properties detection organization, scientific research institution, quality detection department of factory, institutions of higher education teaching and research.
DRX-II-300A Thermal Conductivity Machine
Main features:
1. With real-time measurement and reading devices, it can automatically measure test piece
2. With test piece pressure online real time detection, display and control functions
3. It has fuzzy guarded plate that is hot plate temperature tracking and precision control technology
4. Self-calibration function of the thermal conductivity and heating power
5. Compensation for test piece’s thickness and the average temperature
6. Windows automatic thermal analysis measuring software system, data analysis and reporting output function.
DRX-II-300A Thermal Conductivity Machine
المعلمات التقنية:
| نطاق درجة الحرارة | 0 ~ 100 درجه مئوية, cold plate to 0 درجه مئوية |
| Cooling system | external water bath |
| Hot plate temperature | Room temperature~100℃ |
| دقة التحكم في درجة الحرارة | 0.1درجه مئوية |
| حجم العينة(أقصى) | 300×300×(5-50) mm/ |
| Thermal resistance range | 0.1 ~ 8.0 م2K/W |
| Coefficient of thermal conductivity range | 0.01 ~ 2.3W/m K |
| التكرار | ±1% |
| دقة | ±3% |
| حالة بيئية | room temperature: 18~25℃, رطوبة: 20~80%RH |
| Data result analysis | Automatic thermal analysis software, Windows 10/7/XP , إنجليزي
/or with embedded touch screen control system (customized by user) |
| Supply voltage | AC 220V±5%, power: 3.0كيلووات |















