Acquisition Condition Selection

Acquisition Condition Selection

Choosing optimal acquisition conditions for XRF analysis is a complex and critical part of the art. Selecting the proper acquisition conditions can mean the difference between measuring an element at PPM levels, or not seeing it at all. There are two fundamental principles that must be met to achieve optimal analysis conditions

.أ) There must be a significant source peak above absorption edge energy of the element of interest. This may be either the K or L edge depending on which one is within the measurable range of the instrument with the preference usually going to K line measurements. The closer the source energy is to the absorption edge, the higher the intensity and sensitivity (counts/sec/PPM) will be for the element of interest. The ideal energy would be precisely at the absorption edge energy, but that is usually not possible.

ب) The other fundamental principle is that the background x-rays within the element of interest region should be reduced as much as practical.The difficulty is that these two principles work in opposition to each other, as the best sensitivity is often achieve when the background is highest, and the background is lowest when the sensitivity is worst. Add to this that the best theoretical detection limits are achieved when the sensitivity is highest, while the net count rate extraction, matrix corrections, and long-term analytical stability are best when the background is lowest. Optimal analytical performance is achieved by finding the best compromise between these two principles, given the instrument hardware.

Acquisition Condition Selection

Elements of Interest

The first step to figuring out how to set up an analysis is to determine the elements of interest. This seems obvious, and sometimes it is, but other times it is not. If a sample has never been analyzed for every conceivable element, the odds are high that it contains something that the people producing it are not aware of. Some samples come into a laboratory as complete unknowns. If a sample is not well characterized it is a good idea to perform a qualitative examination of the material using three or more acquisition conditions designed to cover high, medium, and low energy ranges. Qualitative acquisition conditions will be covered below.

Source Selection

IsotopesIsotopes are the simplest source to configure. The most common isotopes are in the table below. Select a source that emits x-rays that are closest to and immediately above the absorption edge energy for the element of interest. To avoid problems with high background the element of interest peak should be at least 2-3 times the FWHM detector resolution away from the source peak.

Isotope Fe-55 Cm-244 Cd-109 Am-241 Co-57

Energy (keV) 5.9 14.3, 18.3 22, 88 59.5 122

Elements (K-lines) آل – V Ti-Br Fe-Mo Ru-Er Ba – ش

Elements (L-lines) Br-II- Pb Yb-Pu None none

2

Sources Selection-X-ray tubesX-ray tube selection is often done by the manufacturer without much input from the customer, but there are some selection rules that are useful. X-ray tubes emit a broad bremstrahlung spectrum that spans from 0 to X KeV where X is the kV potential across the x-ray tube. The peak intensity in the Bremstrahlung spectrum is at roughly 1/2 the maximum energy. The x-ray tube also emits line energies that are characteristic of the target element, so target selection is usually based on selecting a target that will provide optimal excitation for the most important elements of interest. Or, alternatively the selection is based on having a line energy that doesn’t increase the background in the region of any important element. Since the amount of x-ray flux is proportional to the atomic number of the target element, targets such as W are also selected on the basis of having the highest total flux. Targets can be any of several high melting point metals. Common target choices include, Sc, ل, سجل تجاري, الحديد, شركة, في, شركة, ي, زر, شهر, ر.س, Pd, اي جي, دبليو, and Pt.

Multiple target tubes have been used in XRF instrument, and they allow the operator to select the target on a per element basis, using the rule that the target with a line energy immediately above the absorption edge is the one selected, providing it is at least 2-3 times the FWHM detector resolution away from element line. If there is no target available with an emission line above the element of interest select the highest atomic number target available, to maximize the total x-ray flux from the tube.Source Selection – X-ray Tube FiltersFilters can be placed in the x-ray path between the x-ray tube and the sample in order to modify the shape of the source spectrum.

Filters can be made of any element that can be formed into a stable solid or film. They are usually metal or plastic, although plastic filters deteriorate when bombarded by x-rays.

The key to the function of filters is the filter elements absorption edge energy. The filter readily absorbs source x-rays immediately above the absorption edge while those below the absorption edge are transmitted. Very high energy x-rays are also transmitted. This produces a low background valley immediately above the filter’s absorption edge that is great for analyzing elements in the energy range beginning 2 FWHM detector resolution widths above the absorption edge.

Filters also fluoresce their own characteristic line energies, which combine with the Bremstrahlung that is transmitted by the filter to for a peak that looks a little like a right triangle. This secondary filter fluorescence peak can be used as a source peak for elements that are about 3 atomic numbers or more less than the filter element.Multilayer filters are also used. The filter nearest the x-ray tube has an absorption edge above the element of interest and a second filter place on the sample side has an absorption edge below the filter. This makes the source peak less triangular and more monochromatic looking while reducing the background. This is the optimal filter method for exciting one element.

Another type of filter is a neutral absorbing filter such as aluminum or plastic. These filters are intended to filter lower energy source x-rays in order to reduce the background in the region of the element of interest. A thin neutral density filter may be useful for measuring elements like S, or P with an Rh, Pd, or Ag target x-ray tube, while thicker Al filters can eliminate these target peaks entirely creating a source that is good for analyzing x-rays between 2 و 10 keV.

Source Selection – Secondary Targets

The theory behind a secondary target is that the characteristic line energies of the targetmaterial are used to excite the sample. The source points at the target rather than thesample and then the target is oriented so that fluorescence from it shines on the sample.Secondary targets produce a highly monochromatic peak with very low background. Insome instruments the target may be oriented orthogonally with respect to the sample.Since the bremstrahlung is polarized when scattered off a surface, the orthogonalarrangement minimizes scatter at the detector. In either case there is very little energy xDray intensity above the characteristic line energy of the target element.above the element of interest absorption edge energy provided it is not so close that theySelecting the optimal target is simply a matter of choosing the lowest energy target that isCompton scatter doesn’t overlap the region of interest. In most high-resolutioninstruments the element of interest energy should be less than or equal to the element twoatomic numbers below the target element.

Acquisition Condition Selection

Barkla Scatterer

There is another secondary target called a Barkla scatterer that scatters the entirecontinuum rather than relying on producing characteristic line energies. These are almostalways used with polarizing optics, since they would otherwise produce very highbackground intensities. Different scatterers are more efficient at different energies,generally increasing with atomic number. Materials like carbon or boron carbide are usedto excite elements below 5 kV while alumina is used for higher energies.Source

Selection -Overview

Depending on the x-ray tube, x-ray tube filters, and secondary target options in a particular analyzer there may be several options for a given element. So now the question becomes, what is the best choice. Because filters and secondary targets perform differently in deferent energy regimes it is useful to split things up into ranges.Low Energy (<3 keV)

In this range filters and targets do not perform well, so direct excitation using L-lines from Mo, ر.س, Pd and Ag targets will usually give the best performance. A thin neutral density filter can be used to reduce the background somewhat, but it will make the sensitivity worse. In special cases such as measuring Al in concentrated Si, a secondary target helps (Si in this example), because one can be selected that does not excite the major element in the sample.

4

Medium Energy (3-23 keV)

In this regime secondary targets are almost always superior. Metals from Sc to Sn fluoresce well producing monochromatic source peaks with low background. The combination that comes close to rivaling a target in this regime is when the x-ray tube target itself used with a x-ray tube filter of the same element to clean up the background, (EX. Rh tube and filter, or Ti tube and filter).

The next best choices are among x-ray tube filters, and are largely dependant on filter thickness. Thick filters (3 or more half thicknesses*) from Ti-Zn make excellent secondary fluoresors that can rival secondary target performance for lower energy elements using less than a tenth the power. Thin filters (1 ل 3 half thicknesses*) that most analyzers come equipped with are optimal when their absorption edge is below the element of interest energy. (* A half thickness is allows 50% transmission of the filters characteristic K or L x-rays)

The last choices are to use thin filters that have florescence lines above the element of interest, or neutral density filters.

High Energy (>23 كيلو فولت)