BGD 862 Groot Xenon Toetskamer
BGD 862 is a multi-function big xenon light accelerated weathering testers which is equipped with one piece high power (6.5KW)water-cooling xenon lamp, its exposure area arrive 6,500cm2
Powerful functions and reliable test results
◆ Meet with all international standards of xenon test.
◆ Equipped with ATLAS xenon arc lamp , filter and components, ensure to get high and same running parameters. Test results have a good reliability and repeatability comparing with import machines.
◆ Automatic rotating drum-type sample rack with three floors structure maximizes exposure uniformity over all specimens
◆ 6,500cm2 exposure area , can hold different shapes and sizes samples.
◆ Can set the cumulative energy (total irradiance energy)obtained by sample to finish a test procedure.
◆ Advanced cooling system for xenon lamp and intelligent air system.
◆ Chinese or English operation window
BGD 862 Groot Xenon Toetskamer
Can set and control automatically many test parameters
◆ Irradiance energy can be set and accurately controlled(340nm,420nm,300nm~400nm,300nm~800nm). Die “Closed-Loop” control system can monitor and automatically compensate the change of light intensity caused by ageing or other factors.
◆ Working room temperature, BPT(black panel temperature)and BST(black standard temperature)can be set and controlled automatically. With high precision Pt 100 Temperatuur sensor, all these temperatures could be monitored accurately.
◆ Working room relative humidity can be set and controlled automatically. With ultrasonic wave humidifying system, the working room can obtain more stable and uniform humidity distribution.
◆ Can set spray way to sample(front spray or back spray), spray time and spray interval time.
Operation-Menu
Simple, maklik om te gebruik
◆ Colorful touch screen and user-friendly operation interface allow operator to set the test parameters and monitor it easily.
◆ The test procedures can be programmed freely; and up to 16 predetermined procedures can be saved in one time. Every procedure includes up to 10 segments setting data.
◆ With USB interface, operator can export any test parameters with Excel format at any time, convenient to check B-SUN’s running status at any time.
◆ Pure(deionized)water machine is optional. With high purity water, operator can get a more reliable testing result,Chamber has alarm function for monitoring water.
BGD 862 Groot Xenon Toetskamer
Safe and reliable
◆ Many alarm protections:Big irradiance difference, pure water conductivity is over limits, too high temperature of cooling water, too low flow rate of cooling water, over-temperature, heating problem,abnormal lamp power. B-SUN will stop running automatically and show the alarm information in the operation window.
◆ Can select TCP/IP Ethernet interface, the user can tele-control the machine through TCP/IP internet.
Convenient to help customers to solve all problems and do some necessary after service
Working Room
BGD 862 Groot Xenon Toetskamer
Hoof Tegniese Parameters
| bestellings inligting →
Tegniese item ↓ |
BGD 862 Xenon Test Chamber |
| Xenon Lamp | 6.5 KW water cooling long arc xenon lamp |
| Light Filter | Import originally from ATLAS, can simulate indoor or outdoor sunshine spectrum |
| Exposure Area | 6,500 cm2(63-65 pcs standard samples of 15cm×7cm size) |
| Monitoring Method to Irradiance | Four types:340nm,420nm,300nm~400nm,300nm~800nm
Showing at the same time |
| Adjustable Irradiance | See Table B. |
| Lifetime of lamps | 2,000 ure |
| Adjustable Range of BPT | RT~110℃ |
| Adjustable Range of BST | RT~120℃ |
| Adjustable Range of Working room | RT~70℃(Dark) |
| Temperatuur Stability | ±1℃ |
| Temperatuur Uniformity | ≤2℃ |
| Temperature Deviation | ≤ 1℃ |
| Adjustable Humidity | Light:10%-75% Dark:10%-95% |
| Rotate speed of sample | 1 r/min(circle as lamp centre) |
| Spray Funksie | Can set spray continuous time and spray period |
| Water Demands | High purity deionized water(Conductivity<2us/cm) |
| Compressed Air | Clean,oilless compressed air with 0.5MPa pressure,Maks. air supply is near 60L/min. Average air consumption is 10L/min~30L/min(Depends on testing standard) |
| Flow of Deionized Water | 0.2L/min(Add humidity or spray) |
| Power Supply | AC380V±10%,Three-phase four-wire 50Hz;Maks. Current 50A,Maks. Power 9.5KW |
| Algehele grootte | 1,220mm×1,200mm×2,050mm(L×W×H) |
| Net Gewig | 500 KG |
| Structure | |
| Cabinet Material | Working room is made up of high quality stainless steel (SUS316) |
| Chamber Door | Simple door to left with filter window;silicon rubber seal the door edge |
| Voorbeeld Holder | Stainless steel material, its position in the chamber can be adjusted(up, middle or down), Sample crack rotate around the centre of lamp |
| Beheerder | Siemens programmable controller. LCD touch-screen |
| Way of adding humidity | Add humidity by high pressure aerosol |
| Thermometer | Heat insulate type black standard thermometer and black panel thermometer |
| Cooling System | Cooled by compressor and forced-air blast system |
| Security Protections And Error Protections | |
| Xenon Lamp Protection | Over temperature of cooling water,low flow of cooling water,abnormal of lamp power |
| Alarm Function | Big irradiance difference, over-temperature of BPT or BST or Working room, big error of humidity, spraying water shortage, pure water conductivity is over limits.
Operator can set these values and downtime |
| Pure System Protect | Monitor the conductivity of pure water at real-time(Once the conductivity≥2us/cm, the machine will stop and alarm) |
| Protection to Radiation | Xenon lamp will turn off automatically when opening the chamber door, in case xenon lamp radiate operator |
| Installing Environment Requirement | |
| Site Requirement | Net area of installation site ≥12m2;Net height≥2.8m.
Equipped with separate air condition which can ensure a 22℃~30℃ & 30%~80% environment permanently. No strong electromagnetic fields, no high concentration dust,no corrosive gas or flammable materials. |
| Power Source | 380V土10% 3-phases;50Hz;50A power source;Equipped with circuit,ground wire and air circuit breaker. |
| Air Outlet | There should be air outlet(Distance to chamber ≤1.0m,height to ground is 2.5m~2.7m, diameter should be 153mm) |
| Water Supply | Equipped with water supply pipes and valves,Supply water pressure≥2.0kg/cm2;conductivity≤120us/cm |
| Drainage | Draining pipes diameter≥ 50mm,height to ground≤10cm,distance to chamber≤0.5m |
Table A:Main Configuration
| Products | Numbers | Manufacture | Notes |
| Xenon Lamp | 1 pc | Atlas | 6.5KW(Also select other USA brand) |
| Filter | 1 stel | Atlas | Includes five inner filters and one 4.5K/6.5K outer filter(S BORO/S BORO) |
| Control System | 1 stel | SIEMENS | S7-300 PLC+ Siemens analog imput |
| eknterface | 1 pc | MCGS | 10〞 human-computer touch screen |
Table B:Adjustable Range of Irridiance
| Combinating Filters | Adjustable Range of Irridiance(W/㎡) | |||||
| Filters | eknner Filter | Outer Filter | Lamp Power(KW) | 340nm | 420nm | 300-400nm |
| Daylight | Type S Boro | Type S Boro | 2.5~7.5 | 0.25~1.26 | 0.59~2.76 | 29~141 |
| Window Glass | Type S Boro | Soda Lime | 2.5~7.5 | 0.23~1.10 | 0.61~2.76 | 28~129 |
| Extended UV | Quartz | Type S Boro | 2.5~7.5 | 0.29~1.50 | 0.59~2.79 | 32~161 |
BGD 862 Groot Xenon Toetskamer
Stand
ISO 11341《Paints and varnishes-Artificial weathering and exposure to artificial radiation–Exposure to filtered xenon-arc radiation》
ISO 12040 《Graphic Technology – Prints and Printing Inks – Assessment of Light Fastness Using Filtered Xenon Arc Light 》
ISO 16474-1 《Paints and varnishes — Methods of exposure to laboratory light sources — Deel 1: General guidance》
ISO 16474-2 《Paints and varnishes — Methods of exposure to laboratory light sources — Deel 2: Xenon-arc lamps》
ASTM D3451 《Standard Guide for Testing Coating Powders and Powder Coatings》
ASTM D3794《Standard Guide for Testing Coil Coatings》
ASTM D4303 《 Standard Test Methods for Lightfastness of Pigments Used in Artists’ Paints》
ASTM D5010 《Standard Guide for Testing Printing Inks and Related Materials》
ASTM D6577 《Standard Guide for Testing Industrial Protective Coatings》
ASTM D6695 《Standard Practice for Xenon-Arc Exposures of Paint and Related Coatings》
ASTM G151《Standard Practice for Exposing Nonmetallic Materials in Accelerated Test Devices that Use Laboratory Light Sources》
ASTM G155《Practice for Operating Xenon Arc Light Apparatus for Exposure of Non-Metallic Materials》
ISO 4892-1 《Plastics – Methods of Exposure to Laboratory Light Sources – Part 1: General guidance》
ISO 4892-2《Plastics – Methods of Exposure to Laboratory Light Sources – Deel 2: Xenon-arc lamps》
SAE J2412 《Accelerated Exposure of Automotive Interior Trim Components Using a Controlled Irradiance Xenon-Arc Apparatus》
………………………………….
BGD 862 Groot Xenon Toetskamer
Opsionele bykomstighede:
BGD 8158—ATLAS 6.5 kW Xenon Lamp (Water-cooled)
BGD 8159—USA brand 6.5 kW Xenon Lamp (Water-cooled)
BGD 8160— ATLAS 2000 HRS Supply PKG(Includes one BGD 8158 ATLAS 6.5 kW water-cooled lamp,5pcs S65 inner filters,one pc 4.5K/6.5K outer filter)
BGD 8161—6.5KW BGD 2000 HRS Supply PKG(Includes one BGD 8159 USA brand 6.5 kW water-cooled lamp,5pcs W-143 inner filters,one pc W-144 outer filter)
BGD 8171--Purity Water System(100L/h, pure water can less than 0.1µS/cm (input water<200µS/cm))
BGD 8176—Cooling Water Machine(For cooling and cycling pure water, can greatly reduce water consumption)
BGD 8179—Air Compression System(Includes air compressor、air reservoir、freezing dryer、precise filter etc)
BGD 8186—Daylight Filter
BGD 8187—Window Glass Filter
BGD 8192—Metal Frame Air Filters
BGD 8197—Sample shelf
(See Xenon Calibration Radiometer)